A new device for controlled eccentric overloading in training and rehabilitation

Affiliations
Clinical Trial

A new device for controlled eccentric overloading in training and rehabilitation

Anna Frohm et al. Eur J Appl Physiol. 2005 May.

Abstract

The aim of this work was to evaluate a device that allows for eccentric overload to be applied under controlled and safe conditions and it is applicable in exercises commonly used in training and rehabilitation. The machine contains a barbell, which is lowered and raised by a motor, following a predetermined velocity profile. It is capable of handling heavy loads (>500 kg) and is instrumented with a sensor to measure the velocity of the barbell and two scales to measure the vertical component of the ground reaction force. The velocity recordings of the built-in displacement sensor were found to correspond well with those obtained using a motion-capture system. Applying known weights on each scale demonstrated linearity with respect to magnitude and independence regarding location of application. The velocity of the barbell was found to be dependent on the load on the barbell and on the resisting force produced by the individual training in the machine. The combined man-machine reliability was tested using a group of habitually active males (n = 13, 28-55 years) performing squats. Peak voluntary resisting force and position at peak resistance were recorded on two occasions, showing no significant differences and a coefficient of variation of 9% and 22%, respectively. Preliminary observations from training in the machine have been positive both for increasing performance in top athletes and for causing pain relief in patients with diffuse knee problems. The possibility of feedback of the force under each foot makes individual dosage of training load possible, which is valuable, e.g. in rehabilitation of a unilateral injury.

PubMed Disclaimer

References

    1. Eur J Appl Physiol Occup Physiol. 1988;57(5):526-30 - PubMed
    1. Scand J Med Sci Sports. 2001 Aug;11(4):197-206 - PubMed
    1. Sports Med. 1995 Jun;19(6):401-17 - PubMed
    1. Clin Orthop Relat Res. 1986 Jul;(208):65-8 - PubMed
    1. J Orthop Sports Phys Ther. 1997 Oct;26(4):209-13 - PubMed

Publication types