Clinical Trial
. 2005 Feb;83(1):11-9.
doi: 10.1111/j.1600-0420.2005.00408.x.

Three-year follow-up of posterior capsule opacification with two different silicone intraocular lenses

Affiliations
Free article
Clinical Trial

Three-year follow-up of posterior capsule opacification with two different silicone intraocular lenses

Karin Sundelin et al. Acta Ophthalmol Scand. 2005 Feb.
Free article

Abstract

Purpose: To compare posterior capsule opacification (PCO) after cataract surgery with implantation of two silicone intraocular lenses (IOLs) with different designs.

Methods: We carried out a prospective, clinical study of 116 patients randomized to standardized phacoemulsification with implantation of CeeOn Edge (n = 57) or SI40NB (n = 59) IOLs. The follow-up period was 3 years. To evaluate PCO morphologically, digital images were obtained and analysed using evaluation of posterior capsule opacification computer software (epco). The neodymium:YAG (Nd:YAG) capsulotomy rate was recorded.

Results: At 2 and 3 years, the eyes with SI40NB IOLs had significantly more PCO than those with the CeeOn Edge IOLs (p = 0.00014 and p = 0.002). Nine Nd:YAG capsulotomies were performed in the SI40NB group and none in the CeeOn Edge group. This difference was statistically significant (p = 0.003). In some patients a regression of PCO was noticed and confirmed using epco. Statistically less PCO was noted when the capsulorhexis rim was placed so that it covered all 360 degrees of the optic of the IOL.

Conclusions: A clinically and statistically significant difference in PCO development between CeeOn Edge and SI40NB IOLs at 2 and 3 years postoperatively was found. These findings support earlier studies indicating that a sharp edge of the optic is a more important factor in IOL design than IOL material in the prevention of PCO.