Validation of bedside measurements of absolute human renal blood flow by a continuous thermodilution technique
- PMID: 1572191
- DOI: 10.1097/00003246-199205000-00019
Validation of bedside measurements of absolute human renal blood flow by a continuous thermodilution technique
Abstract
Background and methods: There is a clinical need for a system that would allow rapid assessment of renal blood flow in patients with oliguric circulatory shock. A local, continuous thermodilution technique for the measurement of renal venous blood flow, using readily available equipment, was developed. To test the hypothesis that this system would allow measurement of renal blood flow in clinical situations, we compared simultaneous measurements made by the continuous thermodilution technique with measurements of: a) absolute flow measured by volumetric collection in an in vitro flow model; b) renal arterial blood flow measured by electromagnetic flow probe under changing hemodynamic conditions in nine pigs; and c) calculated renal blood flow derived from a clearance technique in 16 patients after cardiac catheterization. The technique utilizes a short-duration, constant infusion of room temperature normal saline into the renal vein via a retrograde thermodilution catheter, with measurement of flow at a thermistor 1 cm back from the tip of the catheter.
Results: The method measured absolute blood flow in an in vitro model, with a correlation coefficient of .99 over blood flows ranging from 55 to 885 mL/min (r2 = .98). There was a .92 correlation coefficient with renal arterial blood flow measured by electromagnetic flow probe in a pig model (r2 = .85), and a .8 correlation with simultaneous measurement of renal blood flow by corrected iodohippurate clearance in humans (r2 = .64). Compared with electromagnetic flow probe measurements, a single measurement by the thermodilution technique would be accurate to within 80 mL/min in 95% of cases. Variability between individual measurements, expressed as the mean of the coefficient of variance for each patient, was 5.5 +/- 3.7%.
Conclusions: This technique is simple to use, requires only venous cannulation and injection of normal saline, and allows rapidly repeatable, immediately available measurements of renal blood flow in a wide range of clinical circumstances, including severe renal impairment or anuria.
Similar articles
-
Bedside estimation of absolute renal blood flow and glomerular filtration rate in the intensive care unit. A validation of two independent methods.Intensive Care Med. 2004 Sep;30(9):1776-82. doi: 10.1007/s00134-004-2380-8. Epub 2004 Jul 28. Intensive Care Med. 2004. PMID: 15375650 Clinical Trial.
-
Evaluation of a new continuous thermodilution cardiac output monitor in critically ill patients: a prospective criterion standard study.Crit Care Med. 1995 May;23(5):860-6. doi: 10.1097/00003246-199505000-00014. Crit Care Med. 1995. PMID: 7736744
-
Direct volumetric blood flow measurement in coronary arteries by thermodilution.J Am Coll Cardiol. 2007 Dec 11;50(24):2294-304. doi: 10.1016/j.jacc.2007.08.047. Epub 2007 Nov 26. J Am Coll Cardiol. 2007. PMID: 18068038
-
Measuring cardiac output: intermittent bolus thermodilution method.Crit Care Nurse. 2000 Apr;20(2):118-20, 122-4. Crit Care Nurse. 2000. PMID: 11873746 Review. No abstract available.
-
Issues in methods and measurement of thermodilution cardiac output.Nurs Res. 1993 Jul-Aug;42(4):228-33. Nurs Res. 1993. PMID: 8337161 Review.
Cited by
-
Bedside estimation of absolute renal blood flow and glomerular filtration rate in the intensive care unit. A validation of two independent methods.Intensive Care Med. 2004 Sep;30(9):1776-82. doi: 10.1007/s00134-004-2380-8. Epub 2004 Jul 28. Intensive Care Med. 2004. PMID: 15375650 Clinical Trial.
-
Increased renal and forearm vasoconstriction in response to exercise after heart transplantation.Br Heart J. 1993 Sep;70(3):247-51. doi: 10.1136/hrt.70.3.247. Br Heart J. 1993. PMID: 8398495 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical