Testing the reliability of a new ultrasonic cardiac output monitor, the USCOM, by using aortic flowprobes in anesthetized dogs
- PMID: 15728064
- DOI: 10.1213/01.ANE.0000144774.42408.05
Testing the reliability of a new ultrasonic cardiac output monitor, the USCOM, by using aortic flowprobes in anesthetized dogs
Abstract
We have used an animal model to test the reliability of a new portable continuous-wave Doppler ultrasonic cardiac output monitor, the USCOM. In six anesthetized dogs, cardiac output was measured with a high-precision transit time ultrasonic flowprobe placed on the ascending aorta. The dogs' cardiac output was increased with a dopamine infusion (0-15 microg x kg(-1) x min(-1)). Simultaneous flowprobe and USCOM cardiac output measurements were made. Up to 64 pairs of readings were collected from each dog. Data were compared by using the Bland and Altman plot method and Lin's concordance correlation coefficient. A total of 319 sets of paired readings were collected. The mean (+/-sd) cardiac output was 2.62 +/- 1.04 L/min, and readings ranged from 0.79 to 5.73 L/min. The mean bias between the 2 sets of readings was -0.0l L/min, with limits of agreement (95% confidence intervals) of -0.34 to 0.31 L/min. This represents a +/-13% error. In five of six dogs, there was a high degree of concordance, or agreement, between the 2 methods, with coefficients >0.9. The USCOM provided reliable measurements of cardiac output over a wide range of values. Clinical trials are needed to validate the device in humans.
Similar articles
-
Ultrasonic cardiac output monitor provides accurate measurement of cardiac output in recipients after liver transplantation.Acta Anaesthesiol Taiwan. 2008 Dec;46(4):171-7. doi: 10.1016/S1875-4597(09)60005-9. Acta Anaesthesiol Taiwan. 2008. PMID: 19097964
-
Reliability of a new ultrasonic cardiac output monitor in recipients of living donor liver transplantation.Liver Transpl. 2008 Jul;14(7):1029-37. doi: 10.1002/lt.21461. Liver Transpl. 2008. PMID: 18581505 Clinical Trial.
-
USCOM (Ultrasonic Cardiac Output Monitors) lacks agreement with thermodilution cardiac output and transoesophageal echocardiography valve measurements.Anaesth Intensive Care. 2007 Dec;35(6):903-10. doi: 10.1177/0310057X0703500608. Anaesth Intensive Care. 2007. PMID: 18084981 Clinical Trial.
-
A meta-analysis of the accuracy and precision of the ultrasonic cardiac output monitor (USCOM).Anaesthesia. 2012 Nov;67(11):1266-71. doi: 10.1111/j.1365-2044.2012.07311.x. Epub 2012 Aug 29. Anaesthesia. 2012. PMID: 22928650 Review.
-
Noninvasive cardiac output measurement in heart failure subjects on circulatory support.Anesth Analg. 2009 Mar;108(3):881-6. doi: 10.1213/ane.0b013e318193174b. Anesth Analg. 2009. PMID: 19224797 Review.
Cited by
-
The effect of head up tilting on bioreactance cardiac output and stroke volume readings using suprasternal transcutaneous Doppler as a control in healthy young adults.J Clin Monit Comput. 2016 Oct;30(5):519-26. doi: 10.1007/s10877-016-9835-7. Epub 2016 Feb 12. J Clin Monit Comput. 2016. PMID: 26873749
-
In vitro evaluation of an ultrasonic cardiac output monitoring (USCOM) device.J Clin Monit Comput. 2016 Feb;30(1):69-75. doi: 10.1007/s10877-015-9685-8. Epub 2015 Mar 7. J Clin Monit Comput. 2016. PMID: 25749977
-
Cardiac Output Measurements in Septic Patients: Comparing the Accuracy of USCOM to PiCCO.Crit Care Res Pract. 2012;2012:270631. doi: 10.1155/2012/270631. Epub 2011 Nov 29. Crit Care Res Pract. 2012. PMID: 22191019 Free PMC article.
-
Stroke Volume Monitoring: Novel Continuous Wave Doppler Parameters, Algorithms and Advanced Noninvasive Haemodynamic Concepts.Curr Anesthesiol Rep. 2017;7(4):387-398. doi: 10.1007/s40140-017-0235-4. Epub 2017 Nov 13. Curr Anesthesiol Rep. 2017. PMID: 29200974 Free PMC article. Review.
-
Ultrasound Cardiac Output Monitor (USCOM™) Measurements Prove Unreliable Compared to Cardiac Magnetic Resonance Imaging in Adolescents with Cardiac Disease.Pediatr Cardiol. 2021 Mar;42(3):692-699. doi: 10.1007/s00246-020-02531-8. Epub 2021 Jan 4. Pediatr Cardiol. 2021. PMID: 33394105
References
-
- Connors AF, Speroff T, Dawson NV, et al. The effectiveness of right heart catheterization in the initial care of critically ill patients. JAMA 1996;276:889–97.
-
- Van De Water JM, Miller TW, Vogel RL, et al. Impedance cardiography: the next vital sign technology? Chest 2003;123:2028–33.
-
- Clarke DE, Raffin TA. Thoracic electrical bioimpedance measurement of cardiac output: not ready for prime time. Crit Care Med 1993;21:1111–2.
-
- Weil MH. Electrical bioimpedance for noninvasive measurement of cardiac output. Crit Care Med 1997;25:1455.
-
- Perrino AC Jr, Fleming J, LaMantia KR. Transesophageal Doppler ultrasonography: evidence for improved cardiac output monitoring. Anesth Analg 1990;71:651–7.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous