Comparative Study
. 1992 Apr;30(4):845-53.
doi: 10.1128/jcm.30.4.845-853.1992.

Comparison of three nonradioisotopic polymerase chain reaction-based methods for detection of human immunodeficiency virus type 1

Affiliations
Comparative Study

Comparison of three nonradioisotopic polymerase chain reaction-based methods for detection of human immunodeficiency virus type 1

A J Whetsell et al. J Clin Microbiol. 1992 Apr.

Abstract

Three nonradioisotopic polymerase chain reaction (PCR)-based detection techniques were evaluated for sensitivity and specificity in detecting human immunodeficiency virus type 1 (HIV-1) proviral DNA in peripheral blood mononuclear cells. The Roche prototype HIV-1 PCR assay, the Du Pont enzyme-linked oligonucleotide sandwich assay (ELOSA), and the Gen-Probe hybridization protection assay (HPA) were compared with a standard radioisotopic oligonucleotide solution hybridization (OSH) technique. A panel of 111 well-characterized clinical samples that included peripheral blood mononuclear cells from 48 healthy, low-risk, HIV-1 antibody-negative subjects, 24 antibody-positive subjects with stable CD4 counts of less than 200/mm3, and 39 antibody-positive subjects with stable CD4 counts of greater than 800/mm3 were studied. Each method demonstrated good specificity, ranging between 96 and 100%; those of the OSH and ELOSA (Du Pont) were 100%, those of the HPA (Gen-Probe) were 100% with one probe and 96% with the other probe, and that of the HIV-1 PCR assay (Roche) was 96%. Sensitivities ranged from 96 to 100% for the low-CD4-count group, with the OSH, the HIV-1 PCR assay (Roche), and the HPA (Gen-Probe) all attaining a sensitivity of 100%. For the high-CD4-count group, sensitivities ranged from 69 to 97%, with the OSH attaining a sensitivity of 97% and the HPA attaining sensitivities of 97% with one probe and 95% with the other probe. These data indicate that the nonradioisotopic techniques are sensitive and specific for the detection of HIV-1 proviral DNA in clinical samples.

PubMed Disclaimer

Similar articles

Cited by

References

    1. J Acquir Immune Defic Syndr. 1991;4(3):277-83 - PubMed
    1. AIDS Res Hum Retroviruses. 1990 Nov;6(11):1323-9 - PubMed
    1. N Engl J Med. 1988 Oct 13;319(15):961-4 - PubMed
    1. Science. 1988 Jan 29;239(4839):487-91 - PubMed
    1. Lancet. 1989 Sep 16;2(8664):637-40 - PubMed

Publication types