Home use of the GlucoWatch G2 biographer in children with diabetes
- PMID: 15741369
- DOI: 10.1542/peds.2004-0820
Home use of the GlucoWatch G2 biographer in children with diabetes
Abstract
Objective: To evaluate usability, accuracy, and hypoglycemia detection of the GlucoWatch G2 Biographer (GW2B) in children aged 1 to 17 years.
Methods: After a 15-hour study of device accuracy, 46 children (15 <7 years, 31 > or =7 years) with type 1 diabetes were enrolled for an extended-wear outcome study: 2 daytime and 2 nighttime 15-hr wear periods each week and blood glucose monitoring 4 times daily for 3 months.
Results: A total of 531 paired GW2B/meter readings were available for accuracy assessment. The correlation coefficients were 0.58 and 0.74 (ages <7 and > or =7 years, respectively). There was no significant change in hemoglobin A1C)or weight-adjusted insulin dose at 3 months after biographer use. Forty-two episodes of hypoglycemia were detected by the GW2B, 33 of which were confirmed by blood glucose meters. Sensitivity and specificity of audible low-glucose alerts were approximately 79% and 83%, respectively. No significant side effects were reported.
Conclusion: The GW2B is usable and safe in children who are <7 years or older in the home setting. The GW2B can detect asymptomatic nocturnal hypoglycemia in younger children.
Similar articles
-
Accuracy of the GlucoWatch G2 Biographer and the continuous glucose monitoring system during hypoglycemia: experience of the Diabetes Research in Children Network.Diabetes Care. 2004 Mar;27(3):722-6. doi: 10.2337/diacare.27.3.722. Diabetes Care. 2004. PMID: 14988292 Free PMC article.
-
A randomized multicenter trial comparing the GlucoWatch Biographer with standard glucose monitoring in children with type 1 diabetes.Diabetes Care. 2005 May;28(5):1101-6. doi: 10.2337/diacare.28.5.1101. Diabetes Care. 2005. PMID: 15855573 Clinical Trial.
-
The accuracy of the GlucoWatch G2 biographer in children with type 1 diabetes: results of the diabetes research in children network (DirecNet) accuracy study.Diabetes Technol Ther. 2003;5(5):791-800. doi: 10.1089/152091503322526996. Diabetes Technol Ther. 2003. PMID: 14633344 Free PMC article. Clinical Trial.
-
Monitoring glycemia in diabetes.Med Clin North Am. 2015 Jan;99(1):35-45. doi: 10.1016/j.mcna.2014.08.017. Epub 2014 Oct 16. Med Clin North Am. 2015. PMID: 25456642 Review.
-
The GlucoWatch biographer: a frequent automatic and noninvasive glucose monitor.Ann Med. 2000 Dec;32(9):632-41. doi: 10.3109/07853890009002034. Ann Med. 2000. PMID: 11209971 Review.
Cited by
-
Development of sensor system and data analytic framework for non-invasive blood glucose prediction.Sci Rep. 2024 Apr 22;14(1):9206. doi: 10.1038/s41598-024-59744-7. Sci Rep. 2024. PMID: 38649731 Free PMC article.
-
Carbon nanotube composites for glucose biosensor incorporated with reverse iontophoresis function for noninvasive glucose monitoring.Int J Nanomedicine. 2010 May 13;5:343-9. Int J Nanomedicine. 2010. PMID: 20517479 Free PMC article.
-
Non-invasive and minimally invasive glucose monitoring devices: a systematic review and meta-analysis on diagnostic accuracy of hypoglycaemia detection.Syst Rev. 2021 May 10;10(1):145. doi: 10.1186/s13643-021-01644-2. Syst Rev. 2021. PMID: 33971958 Free PMC article.
-
Technological Ecological Momentary Assessment Tools to Study Type 1 Diabetes in Youth: Viewpoint of Methodologies.JMIR Diabetes. 2021 Jun 3;6(2):e27027. doi: 10.2196/27027. JMIR Diabetes. 2021. PMID: 34081017 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical