. 2005 Feb;16(2):153-60.
doi: 10.1007/s10856-005-5893-y.

Dynamic creep and mechanical characteristics of SmartSet GHV bone cement

Affiliations

Dynamic creep and mechanical characteristics of SmartSet GHV bone cement

C Z Liu et al. J Mater Sci Mater Med. 2005 Feb.

Abstract

The restrained dynamic creep behaviour and mechanical properties of SmartSet GHV bone cement have been investigated at both room temperature and body temperature. It was found that the bone cement behaves significant differently at room temperature from that at body temperature. The test temperature had a strong effect on the creep performance of the bone cements with a higher creep rate observed at body temperature at each loading cycle. For both temperatures, two stages of creep were identified with a higher creep rate during early cycling followed by a steady state creep rate. The relationship between creep deformation and loading cycle can be expressed by a Hyperb 1 model. As a visco-elastic material, the sensitivity of bone cement to the temperature change was evident during mechanical testing. Compared to the mechanical strength at room temperature, a decreased value was demonstrated at body temperature. The bending modulus was very sensitive to the change in testing temperature, where a reduction of 52% was recorded. A significant reduction in compressive and bending strength, 31 and 23% were recorded respectively. The effect of temperature on bending strength was less apparent, where only 13% reduction was exhibited at body temperature compared to room temperature.

PubMed Disclaimer

References

    1. J Arthroplasty. 2000 Dec;15(8):1028-35 - PubMed
    1. J Biomed Mater Res. 1997 Feb;34(2):221-6 - PubMed
    1. J Biomed Mater Res. 1987 Jun;21(6):719-30 - PubMed
    1. J Biomed Mater Res. 1984 Jan;18(1):25-37 - PubMed
    1. J Biomed Mater Res. 1976 Nov;10(6):837-45 - PubMed

Publication types

LinkOut - more resources