Application of the No Action Level (NAL) protocol to correct for prostate motion based on electronic portal imaging of implanted markers
- PMID: 15752876
- DOI: 10.1016/j.ijrobp.2004.09.035
Application of the No Action Level (NAL) protocol to correct for prostate motion based on electronic portal imaging of implanted markers
Abstract
Purpose: To evaluate the efficacy of the No Action Level (NAL) off-line correction protocol in the reduction of systematic prostate displacements as determined from electronic portal images (EPI) using implanted markers.
Methods and materials: Four platinum markers, two near the apex and two near the base of the prostate, were implanted for localization purposes in patients who received fractionated high dose rate brachytherapy. During the following course of 25 fractions of external beam radiotherapy, the position of each marker relative to the corresponding position in digitally reconstructed radiographs (DRRs) was measured in EPI in 15 patients for on average 17 fractions per patient. These marker positions yield the composite displacements due to both setup error and internal prostate motion, relative to the planning computed tomography scan. As the NAL protocol is highly effective in reducing systematic errors (recurring each fraction) due to setup inaccuracy alone, we investigated its efficacy in reducing systematic composite displacements. The analysis was performed for the center of mass (COM) of the four markers, as well as for the cranial and caudal markers separately. Furthermore, the impact of prostate rotation on the achieved positioning accuracy was determined.
Results: In case of no setup corrections, the standard deviations of the systematic composite displacements of the COM were 3-4 mm in the craniocaudal and anterior-posterior directions, and 2 mm in the left-right direction. The corresponding SDs of the random displacements (interfraction fluctuations) were 2-3 mm in each direction. When applying a NAL protocol based on three initial treatment fractions, the SDs of the systematic COM displacements were reduced to 1-2 mm. Displacements at the cranial end of the prostate were slightly larger than at the caudal end, and quantitative analysis showed this originates from left-right axis rotations about the prostate apex. Further analysis revealed that significant time trends are present in these prostate rotations. No significant trends were observed for the prostate translations.
Conclusions: The NAL protocol based on marker positions in EPI halved the composite systematic displacements using only three imaged fractions per patient, and thus allowed for a significant reduction of planning margins. Although large rotations of the prostate, and time trends therein, were observed, the net impact on the measured displacements and on the accuracy obtained with NAL was small.
Similar articles
-
On-line aSi portal imaging of implanted fiducial markers for the reduction of interfraction error during conformal radiotherapy of prostate carcinoma.Int J Radiat Oncol Biol Phys. 2004 Sep 1;60(1):329-34. doi: 10.1016/j.ijrobp.2004.03.038. Int J Radiat Oncol Biol Phys. 2004. PMID: 15337572
-
The effect of an endorectal balloon and off-line correction on the interfraction systematic and random prostate position variations: a comparative study.Int J Radiat Oncol Biol Phys. 2005 Jan 1;61(1):278-88. doi: 10.1016/j.ijrobp.2004.09.042. Int J Radiat Oncol Biol Phys. 2005. PMID: 15629621
-
Evaluation of multiple image-based modalities for image-guided radiation therapy (IGRT) of prostate carcinoma: a prospective study.Med Phys. 2013 Apr;40(4):041707. doi: 10.1118/1.4794502. Med Phys. 2013. PMID: 23556877 Clinical Trial.
-
[Quantification of prostate movements during radiotherapy].Cancer Radiother. 2006 Nov;10(6-7):381-7. doi: 10.1016/j.canrad.2006.08.003. Cancer Radiother. 2006. PMID: 17049293 Review. French.
-
Chapman on the Diseases and Displacements of the Uterus, &c.Br Foreign Med Chir Rev. 1874 Jan;53(105):74-77. Br Foreign Med Chir Rev. 1874. PMID: 30163195 Free PMC article. Review. No abstract available.
Cited by
-
Intra-fractional uncertainties in image-guided intensity-modulated radiotherapy (IMRT) of prostate cancer.Strahlenther Onkol. 2008 Dec;184(12):668-73. doi: 10.1007/s00066-008-1875-6. Epub 2008 Dec 24. Strahlenther Onkol. 2008. PMID: 19107348
-
Correction of systematic set-up error in breast and head and neck irradiation through a no-action level (NAL) protocol.Clin Transl Oncol. 2011 Jan;13(1):34-42. doi: 10.1007/s12094-011-0614-0. Clin Transl Oncol. 2011. PMID: 21239353
-
Moderately hypofractionated radiotherapy for localized prostate cancer: long-term outcome using IMRT and volumetric IGRT.Strahlenther Onkol. 2014 Jan;190(1):48-53. doi: 10.1007/s00066-013-0443-x. Epub 2013 Nov 8. Strahlenther Onkol. 2014. PMID: 24196279 Clinical Trial.
-
Individualized margins for prostate patients using a wireless localization and tracking system.J Appl Clin Med Phys. 2011 May 22;12(3):3516. doi: 10.1120/jacmp.v12i3.3516. J Appl Clin Med Phys. 2011. PMID: 21844865 Free PMC article.
-
Pretreatment verification of dose calculation and delivery by means of measurements with PLEXITOM™ phantom.Rep Pract Oncol Radiother. 2013 Feb 4;18(2):95-100. doi: 10.1016/j.rpor.2012.12.007. eCollection 2013. Rep Pract Oncol Radiother. 2013. PMID: 24416540 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials
Miscellaneous