Comparison of two bispectral index algorithms in monitoring sedation in postoperative intensive care patients
- PMID: 15753750
- DOI: 10.1097/01.ccm.0000156291.04287.7f
Comparison of two bispectral index algorithms in monitoring sedation in postoperative intensive care patients
Abstract
Objective: Comparison of two versions of bispectral index (BIS) derived from the electroencephalogram in mechanically ventilated and continuously sedated patients after major abdominal surgery.
Design: Prospective, cohort, observational, unblinded study.
Setting: Surgical intensive care unit of a university hospital.
Patients: Forty-six patients undergoing major abdominal surgery scheduled for postoperative mechanical ventilation and continuous sedation.
Interventions: Patients were continuously sedated using propofol and sufentanil.
Measurements and main results: Electrodes for determination of BIS were placed at the forehead of the patients according to the manufacturer's specifications immediately after arrival to the intensive care unit. The level of consciousness was assessed every 2 hrs by a clinical sedation scale (Ramsay sedation scale, levels 1-6). BIS, electromyographic activity, and signal quality index were recorded simultaneously at the same time as the Ramsay scale for 24 hrs or until patients were ready for extubation. BIS (version 2.10) and BIS XP (version 3.12) were determined at 2-hr intervals. BIS and BIS XP showed a high correlation of readings (rs = .79, p < .01). However, the methods did not yield identical results. Both variables were significantly influenced by electromyographic activity, especially at high values, whereas there was only a weak correlation with body temperature. Some of the deeply sedated patients (Ramsay 5 or 6) had BIS readings >80 (BIS, 16%; BIS XP, 13%; p = not significant).
Conclusions: The newer algorithm BIS XP did not perform better than the previous version BIS in patients after major surgery who were mechanically ventilated and sedated on an intensive care unit. This precludes the use of BIS or BIS XP for distinguishing, among deeply sedated ICU patients, those with and without preserved cerebral electrical activity.
Comment in
-
The Bispectral Index: expanded performance for everyday use in the intensive care unit?Crit Care Med. 2005 Mar;33(3):685-7. doi: 10.1097/01.ccm.0000155994.45687.64. Crit Care Med. 2005. PMID: 15753773 No abstract available.
Similar articles
-
Relative reliability of the auditory evoked potential and Bispectral Index for monitoring sedation level in surgical intensive care patients.Anaesth Intensive Care. 2008 Jul;36(4):553-9. doi: 10.1177/0310057X0803600409. Anaesth Intensive Care. 2008. PMID: 18714625
-
Is the bispectral index appropriate for monitoring the sedation level of mechanically ventilated surgical ICU patients?Intensive Care Med. 2002 Feb;28(2):178-83. doi: 10.1007/s00134-001-1183-4. Epub 2002 Jan 12. Intensive Care Med. 2002. PMID: 11907661
-
Consciousness monitoring in ventilated patients: bispectral EEG monitors arousal not delirium.Intensive Care Med. 2004 Aug;30(8):1537-43. doi: 10.1007/s00134-004-2298-1. Epub 2004 May 4. Intensive Care Med. 2004. PMID: 15127189
-
Bispectral index as a guide for titration of propofol during procedural sedation among children.Pediatrics. 2005 Jun;115(6):1666-74. doi: 10.1542/peds.2004-1979. Pediatrics. 2005. PMID: 15930231 Review.
-
Bispectral index monitoring in the intensive care unit provides more signal than noise.Pharmacotherapy. 2005 May;25(5 Pt 2):19S-27S. doi: 10.1592/phco.2005.25.5_part_2.19s. Pharmacotherapy. 2005. PMID: 15899745 Review.
Cited by
-
Bispectral index during epidural puncture predicts anterograde amnesia in patients given midazolam premedication.J Anesth. 2009;23(3):329-33. doi: 10.1007/s00540-009-0748-1. Epub 2009 Aug 14. J Anesth. 2009. PMID: 19685110
-
An assessment of the validity of spectral entropy as a measure of sedation state in mechanically ventilated critically ill patients.Intensive Care Med. 2008 Feb;34(2):308-15. doi: 10.1007/s00134-007-0858-x. Epub 2007 Sep 27. Intensive Care Med. 2008. PMID: 17898996
-
Prognostic evaluation of bispectral index in patients following cardiopulmonary resuscitation.Exp Ther Med. 2013 Mar;5(3):907-911. doi: 10.3892/etm.2013.884. Epub 2013 Jan 4. Exp Ther Med. 2013. PMID: 23407546 Free PMC article.
-
Bispectral index correlates well with Richmond agitation sedation scale in mechanically ventilated critically ill patients.J Anesth. 2010 Jun;24(3):394-8. doi: 10.1007/s00540-010-0915-4. Epub 2010 Mar 12. J Anesth. 2010. PMID: 20225074
-
Novel insights on association and reactivity of Bispectral Index, frontal electromyogram, and autonomic responses in nociception-sedation monitoring of critical care patients.BMC Anesthesiol. 2022 Nov 15;22(1):353. doi: 10.1186/s12871-022-01864-6. BMC Anesthesiol. 2022. PMID: 36380289 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous