Dynamic contour tonometry versus Goldmann applanation tonometry: a comparative study
- PMID: 15756572
- DOI: 10.1007/s00417-005-1124-y
Dynamic contour tonometry versus Goldmann applanation tonometry: a comparative study
Abstract
Background: Various sources of error, including central corneal thickness (CCT) and structural corneal rigidity, have been proposed for Goldmann applanation tonometry (GAT). The Pascal dynamic contour tonometer (DCT) is a novel device designed for intraocular pressure (IOP) measurements assumed to be largely independent of CCT and corneal curvature. We compared DCT with GAT in eyes with normal corneas of various thickness.
Methods: We prospectively measured IOP using DCT and GAT in random order in 100 eyes of 100 subjects (M:F=46:54; mean age 42+/-19, range 23-88 years).
Results: Mean DCT values were about 1 mm Hg higher than GAT readings (16+/-3 vs 15+/-3 mmHg, p=0.001). Bland-Altman analysis of individual pairs of DCT and GAT measurements revealed a bias of -1.0 mmHg [95% confidence interval (CI): +/-1.2]. Neither GAT nor DCT showed a significant correlation with CCT (533+/-48, range 399-641 microm).
Conclusions: In eyes with normal corneas, DCT allows suitable and reliable IOP measurements which are in good concordance with GAT. Comparison of DCT with intracameral manometry is desirable in the future.
Similar articles
-
Comparisons between Pascal dynamic contour tonometry, the TonoPen, and Goldmann applanation tonometry in patients with glaucoma.Acta Ophthalmol Scand. 2007 May;85(3):272-9. doi: 10.1111/j.1600-0420.2006.00834.x. Acta Ophthalmol Scand. 2007. PMID: 17488456
-
Effects of corneal thickness, corneal curvature, and intraocular pressure level on Goldmann applanation tonometry and dynamic contour tonometry.Ophthalmology. 2007 Jan;114(1):20-6. doi: 10.1016/j.ophtha.2006.06.047. Epub 2006 Oct 27. Ophthalmology. 2007. PMID: 17070592
-
Dynamic contour tonometry (DCT) versus Goldmann applanation tonometry (GAT) - a comparison of agreement and reproducibility.Graefes Arch Clin Exp Ophthalmol. 2007 Jul;245(7):1027-30. doi: 10.1007/s00417-006-0449-5. Epub 2006 Dec 20. Graefes Arch Clin Exp Ophthalmol. 2007. PMID: 17177039
-
The influence of altitude on the differences between Goldmann tonometry and Pascal dynamic contour tonometry: An ecological meta-analysis.Indian J Ophthalmol. 2024 May 1;72(Suppl 3):S398-S403. doi: 10.4103/IJO.IJO_907_23. Epub 2023 Dec 15. Indian J Ophthalmol. 2024. PMID: 38099358 Free PMC article. Review.
-
Finite element simulation of Goldmann tonometry after refractive surgery.Clin Biomech (Bristol). 2020 Jan;71:24-28. doi: 10.1016/j.clinbiomech.2019.09.007. Epub 2019 Oct 5. Clin Biomech (Bristol). 2020. PMID: 31677547
Cited by
-
The effect of contact lens-induced corneal edema on Goldmann applanation tonometry and dynamic contour tonometry.Graefes Arch Clin Exp Ophthalmol. 2009 Mar;247(3):371-5. doi: 10.1007/s00417-008-0954-9. Epub 2008 Oct 9. Graefes Arch Clin Exp Ophthalmol. 2009. PMID: 18843499
-
Goldmann applanation tonometry and dynamic contour tonometry in eyes with elevated intraocular pressure (IOP): comparison in the same eyes after subsequent medical normalization of IOP.Graefes Arch Clin Exp Ophthalmol. 2010 Nov;248(11):1611-6. doi: 10.1007/s00417-010-1462-2. Epub 2010 Jul 22. Graefes Arch Clin Exp Ophthalmol. 2010. PMID: 20652303
-
Distribution of central corneal thickness and its association with ocular parameters in a large central European cohort: the Gutenberg health study.PLoS One. 2013 Aug 1;8(8):e66158. doi: 10.1371/journal.pone.0066158. Print 2013. PLoS One. 2013. PMID: 23936291 Free PMC article.
-
Comparison of Goldmann and Pascal tonometry in relation to corneal hysteresis and central corneal thickness in nonglaucomatous eyes.Clin Ophthalmol. 2011;5:1071-7. doi: 10.2147/OPTH.S23086. Epub 2011 Aug 3. Clin Ophthalmol. 2011. PMID: 21847339 Free PMC article.
-
Correlation of intraocular pressure measured with goldmann and dynamic contour tonometry in normal and glaucomatous eyes.J Glaucoma. 2009 Feb;18(2):119-23. doi: 10.1097/IJG.0b013e31817d23c7. J Glaucoma. 2009. PMID: 19225347 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources