[Clinical results with A new retrograde femoral nail with a radiographical-free proximal locking device. A prospective study with 50 fractures]
- PMID: 15778831
- DOI: 10.1007/s00113-004-0863-2
[Clinical results with A new retrograde femoral nail with a radiographical-free proximal locking device. A prospective study with 50 fractures]
Abstract
Introduction: Conventional retrograde nailing of the femur causes two important disadvantages: the proximal locking of the nail is difficult because of the anatomic conditions and a chondral defect was left into the knee.
Material and methods: After the retrograde implantation the new nail was lead through the greater trochanter. An additional proximal aiming device for proximal interlocking can be fixed. The entrance portal will be sealed by an osteochondral cylinder. 50 cases of femur fractures were selected for the prospective study. We recorded all intraoperative complications and technical difficulties. The cases were followed up for 52 weeks, both clinical and radiology examinations were performed.
Results: The mean follow up was 15.5+/-5 months. All fractures were healed. Knee movement was 125+/-14 degrees. The Leung Score was 84+/-12.6 points; HSS Score was 90+/-9 points. In two cases wound infections were developed. Mal-union was observed in three cases, in two cases nail brake down.
Conclusions: The new retrograde interlocking nail could be used to manage femur fractures successfully. Two aiming devices enable a easy interlocking. Replacement of the osteochondral cylinder into the entry portal reduces cartilage damage.
Similar articles
-
Retrograde femoral interlocking nail in complex fractures.J Orthop Surg (Hong Kong). 2002 Jun;10(1):17-21. doi: 10.1177/230949900201000104. J Orthop Surg (Hong Kong). 2002. PMID: 12401916
-
Cephalomedullary nails in the treatment of high-energy proximal femur fractures in young patients: a prospective, randomized comparison of trochanteric versus piriformis fossa entry portal.J Orthop Trauma. 2006 Apr;20(4):240-6. doi: 10.1097/00005131-200604000-00002. J Orthop Trauma. 2006. PMID: 16721238 Clinical Trial.
-
Cephalomedullary screws as the standard proximal locking screws for nailing femoral shaft fractures.J Orthop Trauma. 2010 Dec;24(12):717-22. doi: 10.1097/BOT.0b013e3181f6bfe8. J Orthop Trauma. 2010. PMID: 21076246
-