Correlation of Narcotrend Index, entropy measures, and spectral parameters with calculated propofol effect-site concentrations during induction of propofol-remifentanil anaesthesia

Affiliations
Clinical Trial

Correlation of Narcotrend Index, entropy measures, and spectral parameters with calculated propofol effect-site concentrations during induction of propofol-remifentanil anaesthesia

Ulrich Grouven et al. J Clin Monit Comput. 2004 Aug.

Abstract

Objective: The aim of this study was to compare the EEG parameter Narcotrend Index with the spectral and entropy-based EEG parameters median frequency, 95% spectral edge frequency, burst-compensated 95% spectral edge frequency, spectral entropy, amplitude entropy, and approximate entropy with regard to their ability to describe cerebral anaesthetic drug effects during induction of propofol-remifentanil anaesthesia.

Methods: Three induction schemes were studied with 10 patients each receiving 2 mg propofol/kg/60s (group 1), 4 mg/kg/120s (group 2), and 4 mg/kg/240s (group 3). The EEG was recorded with the EEG monitor Narcotrend. To analyse the relation between drug effect and EEG parameters, Spearman rank correlation of the different EEG parameters with the calculated propofol effect-site concentration was computed.

Results: In all groups Narcotrend Index showed the highest correlation with the propofol effect-site concentration and the lowest variability of individual correlation values. Furthermore, only the Narcotrend Index showed a monophasic behaviour over the entire time period analysed. In the group of entropy parameters approximate entropy yielded the best results. Among the spectral parameters the burst-compensated 95% spectral edge frequency had the highest correlation with the propofol effect-site concentration. It was markedly higher than for the standard spectral edge frequency. The correlations of median frequency and amplitude entropy with propofol effect-site concentration were the lowest. CONCLUSIONS. Changes in the propofol effect-site concentration during induction of anaesthesia were best described by the multivariate Narcotrend Index compared to conventional spectral EEG parameters and different entropy measures.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Anaesth Intensive Care. 2002 Dec;30(6):817-8 - PubMed
    1. Proc Natl Acad Sci U S A. 1991 Mar 15;88(6):2297-301 - PubMed
    1. Anaesthesia. 1998 Apr;53 Suppl 1:35-41 - PubMed
    1. Anesthesiology. 2003 Jan;98(1):53-7 - PubMed
    1. J Clin Monit Comput. 2002 Feb;17(2):125-34 - PubMed

Publication types

LinkOut - more resources