Ex vivo experiment of saline-enhanced hepatic bipolar radiofrequency ablation with a perfused needle electrode: comparison with conventional monopolar and simultaneous monopolar modes
- PMID: 15789259
- DOI: 10.1007/s00270-004-0177-3
Ex vivo experiment of saline-enhanced hepatic bipolar radiofrequency ablation with a perfused needle electrode: comparison with conventional monopolar and simultaneous monopolar modes
Abstract
The purpose of this study was to validate the saline-enhanced bipolar radiofrequency ablation (RFA) technique using a perfused electrode to increase RF-created coagulation necrosis, to compare that technique with monopolar RFAs and to find appropriate concentrations and volumes of perfused NaCl solution for the bipolar RFA. A total of 90 ablations were performed in explanted bovine livers. In the initial experiments to determine appropriate conditions for bipolar RFA, we created five thermal ablation zones in each condition, with instillations of varied concentrations (0.9-36%) or injection rates (30 mL/hr-120 mL/hr) of NaCl solution. After placement of one or two 16-gauge open-perfused electrodes into bovine livers, the NaCl solution was instilled into the tissue through the electrode. In the second part of the study, 10 ablation zones were created using one or two perfused electrodes for each of five groups under different conditions: a conventional monopolar mode with 0.9% NaCl solution (group A) or with 6% NaCl solution (group B), a simultaneous monopolar mode with 6% NaCl solution (group C) and a bipolar mode with 6% NaCl solution (groups D and E). RF was applied to each electrode for 20 min in groups A, B, C, and E, or for 10 min in group D. During RFA, we measured the tissue temperature 15 mm from the electrode. The temperature changes during the RFA and the dimensions of the ablation zones were compared among the groups. Bipolar RFA created larger short-axis diameters of coagulation necrosis with 6% NaCl solution (35.8 +/- 15 mm) than with 0.9% NaCl solution (17 +/- 9.7 mm) (P < 0.05). However, concentrations of NaCl solution above 6% did not further increase the extent of coagulation necrosis. In addition, bipolar RFA with 6% NaCl solution instillation at a rate of 1.0 mL/min (37.9 +/- 5.4 mm) or 2.0 mL/min (35.6 +/- 9.3 mm) produced larger diameters at the mid-point between the electrodes of the ablated lesion than did 0.5 mL/min (25.8 +/- 9.3 mm) (P < 0.05). The bipolar mode showed a more rapid increase in temperature at the mid-point between the two probes, up to 60 degrees C, than did the monopolar modes (P < 0.05). In addition, the bipolar RFA (group E) treated for 20 min showed a larger value of the short-axis diameter than did the conventional or simultaneous monopolar modes (P < 0.05), and bipolar RFA (group D) treated for 10 min, showed similar results with conventional monopolar modes treated for 20 min (P > 0.05): 31.0 +/- 5.4 mm (group A); 28.8 +/- 3.8 mm (group B); 25.5 +/- 6.4 mm (group C); 32.6 +/- 4.2 mm (group D); 49.4 +/- 5.0 mm (group E). Bipolar RFA with instillation of 6% NaCl solution through an open perfusion system demonstrates better efficacy in creating a larger ablation zone than does conventional or simultaneous monopolar modes at the various times examined. Therefore, hypertonic saline-enhanced bipolar RFA seems to be a promising approach for treating larger liver tumors.
Similar articles
-
Wet radio-frequency ablation using multiple electrodes: comparative study of bipolar versus monopolar modes in the bovine liver.Eur J Radiol. 2005 Jun;54(3):408-17. doi: 10.1016/j.ejrad.2004.06.004. Eur J Radiol. 2005. PMID: 15899344
-
A comparative experimental study of the in-vitro efficiency of hypertonic saline-enhanced hepatic bipolar and monopolar radiofrequency ablation.Korean J Radiol. 2003 Jul-Sep;4(3):163-9. doi: 10.3348/kjr.2003.4.3.163. Korean J Radiol. 2003. PMID: 14530645 Free PMC article.
-
Comparison of wet radiofrequency ablation with dry radiofrequency ablation and radiofrequency ablation using hypertonic saline preinjection: ex vivo bovine liver.Korean J Radiol. 2004 Oct-Dec;5(4):258-65. doi: 10.3348/kjr.2004.5.4.258. Korean J Radiol. 2004. PMID: 15637476 Free PMC article.
-
Needle-based ablation of renal parenchyma using microwave, cryoablation, impedance- and temperature-based monopolar and bipolar radiofrequency, and liquid and gel chemoablation: laboratory studies and review of the literature.J Endourol. 2004 Feb;18(1):83-104. doi: 10.1089/089277904322836749. J Endourol. 2004. PMID: 15006061 Review.
-
Radiofrequency ablation of hepatocellular carcinoma.Minerva Chir. 2002 Apr;57(2):165-76. Minerva Chir. 2002. PMID: 11941292 Review.
Cited by
-
Perfused hypertonic-saline-augmented needle enlarges ablation zones in ex vivo porcine livers.Oncol Lett. 2018 Jan;15(1):972-978. doi: 10.3892/ol.2017.7352. Epub 2017 Nov 6. Oncol Lett. 2018. PMID: 29391896 Free PMC article.
-
Radiofrequency ablation using internally cooled wet electrodes in bipolar mode for the treatment of recurrent hepatocellular carcinoma after locoregional treatment: A randomized prospective comparative study.PLoS One. 2020 Sep 28;15(9):e0239733. doi: 10.1371/journal.pone.0239733. eCollection 2020. PLoS One. 2020. PMID: 32986758 Free PMC article. Clinical Trial.
-
No-Touch Radiofrequency Ablation: A Comparison of Switching Bipolar and Switching Monopolar Ablation in Ex Vivo Bovine Liver.Korean J Radiol. 2017 Mar-Apr;18(2):279-288. doi: 10.3348/kjr.2017.18.2.279. Epub 2017 Feb 7. Korean J Radiol. 2017. PMID: 28246508 Free PMC article.
-
Comparison of Ablation Performance between Octopus Multipurpose Electrode and Conventional Octopus Electrode.Korean J Radiol. 2023 Feb;24(2):86-94. doi: 10.3348/kjr.2022.0600. Korean J Radiol. 2023. PMID: 36725351 Free PMC article.
-
Characterization of the ablation zones produced by three commercially available systems from a single vendor for radiofrequency thermoablation in an ex vivo swine liver model.Vet Med Sci. 2020 Nov;6(4):1041-1048. doi: 10.1002/vms3.319. Epub 2020 Jul 1. Vet Med Sci. 2020. PMID: 32613738 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources