Multiplexed AtheNA multi-lyte immunoassay for ANA screening in autoimmune diseases
- PMID: 15804711
- DOI: 10.1080/08916930400022707
Multiplexed AtheNA multi-lyte immunoassay for ANA screening in autoimmune diseases
Abstract
Background: Multiplexed assays using fluorescence microspheres is an exciting technology with multiple applications including the detection of antinuclear autoantibodies (ANA) and autoantibody profiles. It is a rapid, sensitive and automatic method for simultaneous quantitative detection of several autoantibodies. The aim of our study was to determinate ANA and other autoantibodies to the nine extractable nuclear antigens by the AtheNA Multi-Lyte ANA system and compare the results achieved by this method to the routinely used enzyme immunoassay.
Methods: Four hundred eighteen serum samples were tested utililizing the multiplexed method: 96 healthy donors, 86 requested ANA specimens obtained from routine lab, and 236 samples from patients with known autoimmune diseases (43-scleroderma, 113-systemic lupus erythematosus, 38-Sjogren's syndrome, and 42 rheumatoid arthritis). The ANA and antibodies to nine different analytes (SS/A, SS/B, Sm, RNP, Jo-1, Scl-70, dsDNA, Centromere B and Histone) were tested.
Results: ANA screening by AtheNA system revealed high concordance of 99 and 97.7% with the enzyme immunoassay test in samples obtained from healthy donors and ANA requested samples, respectively. Evaluation of autoimmune disease-related samples for ANA by AtheNA technology also confirmed a high rate of concordance of 92-97.7% and correlated with the enzyme immunoassay. Positive discrepant results were found for Scl-70 specificity in 12.7% of SLE specimens by AtheNA technology, while all tested sera were negative for this antibody by enzyme immunoassay. Negative discrepant results were observed by the AtheNA system for anti-dsDNA. The sera (15 randomly obtained samples from SLE patients) were positive for anti-dsDNA in 50% of samples in Farr assay and 55% in enzyme immunoassay, respectively.
Conclusion: We suggest that the AtheNA technology may be a useful diagnostic tool for ANA screening. Additional investigations are required to compare an analytic performance between AtheNA and routine methods in determination of the individual autoantibody profile.
Similar articles
-
Clinical value of multiplexed bead-based immunoassays for detection of autoantibodies to nuclear antigens.Clin Vaccine Immunol. 2007 May;14(5):505-9. doi: 10.1128/CVI.00034-07. Epub 2007 Mar 21. Clin Vaccine Immunol. 2007. PMID: 17376860 Free PMC article. Clinical Trial.
-
Evaluation of the BioPlex 2200 ANA screen for the detection of antinuclear antibodies and comparison with conventional methods.Ann N Y Acad Sci. 2007 Aug;1109:245-55. doi: 10.1196/annals.1398.030. Ann N Y Acad Sci. 2007. PMID: 17785313
-
Effective use of autoantibody tests in the diagnosis of systemic autoimmune disease.Ann N Y Acad Sci. 2005 Jun;1050:217-28. doi: 10.1196/annals.1313.023. Ann N Y Acad Sci. 2005. PMID: 16014537 Review.
-
Evaluation of the BioPlex 2200 ANA screen: analysis of 510 healthy subjects: incidence of natural/predictive autoantibodies.Ann N Y Acad Sci. 2005 Jun;1050:380-8. doi: 10.1196/annals.1313.120. Ann N Y Acad Sci. 2005. PMID: 16014555
-
Diagnostic value of screening enzyme immunoassays compared to indirect immunofluorescence for anti-nuclear antibodies in patients with systemic rheumatic diseases: A systematic review and meta-analysis.Semin Arthritis Rheum. 2018 Oct;48(2):334-342. doi: 10.1016/j.semarthrit.2018.01.011. Epub 2018 Mar 31. Semin Arthritis Rheum. 2018. PMID: 29609799
Cited by
-
Clinical comparison of QUANTA Flash dsDNA chemiluminescent immunoassay with four current assays for the detection of anti-dsDNA autoantibodies.J Immunol Res. 2015;2015:902821. doi: 10.1155/2015/902821. Epub 2015 Jan 5. J Immunol Res. 2015. PMID: 25759849 Free PMC article.
-
Current concepts and future directions for the assessment of autoantibodies to cellular antigens referred to as anti-nuclear antibodies.J Immunol Res. 2014;2014:315179. doi: 10.1155/2014/315179. Epub 2014 Apr 27. J Immunol Res. 2014. PMID: 24868563 Free PMC article. Review.
-
Analytical variability in the determination of anti-double-stranded DNA antibodies: the strong need of a better definition of the old and new tests.Immunol Res. 2018 Jun;66(3):340-347. doi: 10.1007/s12026-018-8992-9. Immunol Res. 2018. PMID: 29623613
-
Antinuclear antibodies and their detection methods in diagnosis of connective tissue diseases: a journey revisited.Diagn Pathol. 2009 Jan 2;4:1. doi: 10.1186/1746-1596-4-1. Diagn Pathol. 2009. PMID: 19121207 Free PMC article.
-
A comparison of anti-nuclear antibody quantification using automated enzyme immunoassays and immunofluorescence assays.Autoimmune Dis. 2014;2014:534759. doi: 10.1155/2014/534759. Epub 2014 Jan 28. Autoimmune Dis. 2014. PMID: 24592328 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials
Miscellaneous