Coil thickness and packing of cerebral aneurysms: a comparative study of two types of coils

Affiliations
Comparative Study

Coil thickness and packing of cerebral aneurysms: a comparative study of two types of coils

Marjan J Slob et al. AJNR Am J Neuroradiol. 2005 Apr.

Abstract

Background and purpose: In coiling cerebral aneurysms, high packing prevents reopening over time. The purpose of this study was to compare packing of cerebral aneurysms treated with two types of coils with different wire thickness and different shapes.

Methods: Packing, defined as the ratio between the volume of inserted coils and the volume of an aneurysm, was calculated for 144 cerebral aneurysms treated in 130 patients. Seventy-two aneurysms were treated with predominantly helical-shaped coils of 0.010-inch-diameter wire, and 72 aneurysms were treated with predominantly complex-shaped coils of 0.012-inch-diameter wire. Aneurysm volume was assessed from three-dimensional angiography. Aneurysm packing, inserted coil lengths, and numbers of coils were compared for both types of coils.

Results: Mean packing was significantly higher (absolute value, 6.4%; relative value, 26.6%; P < .0001) in aneurysms coiled with 0.012-inch-diameter coils than aneurysms coiled with 0.010-inch-diameter coils. Inserted coil length per cubic millimeter of aneurysmal volume was equal for both types of coils.

Conclusion: Use of complex-shaped coils with a wire diameter of 0.012 inch to coil cerebral aneurysms results in significantly better packing than is achieved with helical coils of 0.010-inch-diameter wire.

PubMed Disclaimer

Figures

F<sc>ig</sc> 1.
Fig 1.
Graph showing the relation between aneurysm volume and packing. Aneurysm volume is a logarithmic scale. Packing is lower with larger aneurysm volumes for both types of coils. Difference in position of trend lines reflects difference in packing.

References

    1. Sluzewski M, van Rooij WJ, Rinkel GJ, Wijnalda D. Endovascular treatment of ruptured intracranial aneurysms with detachable coils: long-term clinical and serial angiographic results. Radiology 2003;227:720–724 - PubMed
    1. Raymond J, Guilbert F, Weill A, et al. Long-term angiographic recurrences after selective endovascular treatment of aneurysms with detachable coils. Stroke 2003;34:1398–1403 - PubMed
    1. Sluzewski M, van Rooij WJ, Slob MJ, et al. Relation between aneurysm volume, packing and compaction in 145 cerebral aneurysms treated with coils. Radiology 2004;231:653–658 - PubMed
    1. Kawanabe Y, Sadato A, Taki W, Hashimoto N. Endovascular occlusion of intracranial aneurysms with Guglielmi detachable coils: correlation between coil packing density and compaction. Acta Neurochir (Wien) 2001;143:451. - PubMed
    1. Tamatani S, Ito Y, Abe H, et al. Evaluation of the stability of aneurysms after embolization using detachable coils: correlation between stability of aneurysms and embolized volume of aneurysms. AJNR Am J Neuroradiol 2002;23:762–767 - PMC - PubMed