. 2005 Apr 1;61(5):1603-6.
doi: 10.1016/j.ijrobp.2004.12.034.

In vivo measurements with MOSFET detectors in oropharynx and nasopharynx intensity-modulated radiation therapy

Affiliations

In vivo measurements with MOSFET detectors in oropharynx and nasopharynx intensity-modulated radiation therapy

Serge Marcié et al. Int J Radiat Oncol Biol Phys. .

Abstract

Purpose: To evaluate the feasibility of in vivo measurements with metal oxide semiconductor field effect transistor (MOSFET) dosimeters for oropharynx and nasopharynx intensity-modulated radiation therapy (IMRT).

Methods and materials: During a 1-year period, in vivo measurements of the dose delivered to one or two points of the oral cavity by IMRT were obtained with MOSFET dosimeters. Measurements were obtained during each session of 48 treatment plans for 21 patients, all of whom were fitted with a custom-made mouth plate. Calculated and measured values were compared.

Results: A total of 344 and 452 measurements were performed for the right and left sides, respectively, of the oral cavity. Seventy percent of the discrepancies between calculated and measured values were within +/-5%. Uncertainties were due to interfraction patient positions, intrafraction patient movements, and interfraction MOSFET positions. Nevertheless, the discrepancies between the measured and calculated means were within +/-5% for 92% and 95% of the right and left sides, respectively. Comparison of these discrepancies and the discrepancies between calculated values and measurements made on a phantom revealed that all differences were within +/-5%.

Conclusion: Our experience demonstrates the feasibility of in vivo measurements with MOSFET dosimeters for oropharynx and nasopharynx IMRT.