Abstract
A validation study was performed on the MESAM 4, a digital recording device developed to monitor oxygen saturation, heart rate (HR), snoring, and body position in order to screen subjects for obstructive sleep apnea syndrome (OSAS). MESAM 4 recordings were scored with the computer-based automatic scoring system provided with the equipment. Nocturnal polysomnography (PSG) and MESAM 4 recordings were run simultaneously on 56 subjects presenting with any type of sleep complaint, including those secondary to OSAS. Patients were assigned to one room by hospital administration and were monitored consecutively. The polygraphic equipment and MESAM 4 equipment were placed on the subjects by separate teams. Records of PSG and MESAM 4 were analyzed in double-blind fashion. With the MESAM 4 computerized analysis, three indices based on SaO2 (ODI), on heart rate (HVI), and on snoring (ISI) were obtained, and the number of abnormal respiratory events occurring during the time selected for analysis (TAT) were determined. Polysomnographic records were scored by 30-s epochs following the American Sleep Disorders Association standards for sleep states and stages and for sleep-related events, including sleep apneas, hypopneas, and periodic leg movements. Following independent scoring, 26 subjects were identified with OSAS by PSG, while MESAM 4 identified 25 subjects with OSAS using oxygen algorithm; all had a respiratory disturbance index greater than or equal to 10 with PSG. Results of each polysomnogram and each MESAM 4 analysis were compared. With the polysomnogram used as a standard, the degree of error for each variable with the MESAM 4 was calculated. Specificity and sensitivity of the most accurate index of the MESAM 4, the ODI, were 97 percent and 92 percent, respectively. The other two indices, HVI and ISI, were less accurate: specificity and sensitivity were 32 percent and 58 percent for HVI and 27 percent and 96 percent for ISI. Nevertheless, a combination of all three indices (ODI, HVI, ISI) would have prevented the two false-positive cases we observed. The results of this validation study show that MESAM 4 can be helpful to general practitioners, clinicians, and epidemiologists as a low-cost screening device for subjects with OSAS and habitual snoring.
Similar articles
-
Investigations of an automatic screening device (MESAM) for obstructive sleep apnoea.Eur Respir J. 1990 Jul;3(7):823-9. Eur Respir J. 1990. PMID: 2261972
-
Hand-scoring of MESAM 4 recordings is more accurate than automatic analysis in screening for obstructive sleep apnoea.Eur Respir J. 1994 Oct;7(10):1771-5. doi: 10.1183/09031936.94.07101771. Eur Respir J. 1994. PMID: 7828683
-
Quantification of sleep disordered breathing by computerized analysis of oximetry, heart rate and snoring.Eur Respir J. 1991 Jun;4(6):655-9. Eur Respir J. 1991. PMID: 1889491
-
Epidemiology of obstructive sleep apnea syndrome.Sleep. 1992 Dec;15(6 Suppl):S1-4. doi: 10.1093/sleep/15.suppl_6.s1. Sleep. 1992. PMID: 1470800 Review.
-
Monitoring at home.Lung. 1990;168 Suppl:927-32. doi: 10.1007/BF02718229. Lung. 1990. PMID: 2117213 Review.
Cited by
-
Use of home sleep studies for diagnosis of the sleep apnoea/hypopnoea syndrome.Thorax. 1997 Dec;52(12):1068-73. doi: 10.1136/thx.52.12.1068. Thorax. 1997. PMID: 9516901 Free PMC article.
-
Nocturnal hypoxemia and oxygen desaturation events in neutropenic patients with sepsis or pneumonia.Intensive Care Med. 1996 Feb;22(2):174. doi: 10.1007/BF01720726. Intensive Care Med. 1996. PMID: 8857126 No abstract available.
-
Obstructive sleep apnea in New Zealand adults: prevalence and risk factors among Māori and non-Māori.Sleep. 2009 Jul;32(7):949-56. doi: 10.1093/sleep/32.7.949. Sleep. 2009. PMID: 19639758 Free PMC article.
-
Validation of the Somnolyzer 24×7 automatic scoring system in children with suspected obstructive sleep apnea.Front Med (Lausanne). 2025 Jun 18;12:1617530. doi: 10.3389/fmed.2025.1617530. eCollection 2025. Front Med (Lausanne). 2025. PMID: 40606446 Free PMC article.
-
Is sleep apnea an independent risk factor for prevalent and incident diabetes in the Busselton Health Study?J Clin Sleep Med. 2009 Feb 15;5(1):15-20. J Clin Sleep Med. 2009. PMID: 19317376 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical