. 2005 Apr;28(4):316-23.
doi: 10.1111/j.1540-8159.2005.09477.x.

Use of advanced mapping systems to guide ablation in complex cases: experience with noncontact mapping and electroanatomic mapping systems

Affiliations

Use of advanced mapping systems to guide ablation in complex cases: experience with noncontact mapping and electroanatomic mapping systems

Osnat T Gurevitz et al. Pacing Clin Electrophysiol. 2005 Apr.

Abstract

Objective: This report describes our experience with noncontact mapping and electroanatomic mapping in complex ablations, which are defined as ablations done after failure of conventional ablation.

Material and methods: Patients were included (N = 68; 49% with structural heart disease) in whom previous ablation failed and in whom a second procedure was done with advanced mapping. Non-contact mapping was used in 17 patients, electroanatomic mapping in 36, and both noncontact and electroanatomic mapping in 15. Arrhythmias included focal atrial tachycardia (n = 16), reentrant atrial tachycardia (n = 14), right ventricular outflow tachycardia (n = 10), post-myocardial infarction ventricular tachycardia (n = 9), and others (n = 19).

Results: Acute success at the second ablation was achieved in 79% of patients. At 20 +/- 9 months after the procedure, 69% of these patients reported having significantly fewer symptoms than before the second ablation, and 51% were free of symptoms. Only 16% were using antiarrhythmic medications. Complications included a small pericardial effusion in two patients, hypotension in one patient, and a femoral pseudoaneurysm in another.

Conclusions: Advanced mapping is a useful and safe adjunct for catheter ablation after ablation has failed in patients with complex substrate.

PubMed Disclaimer