The Dinamap 1846SX automated blood pressure recorder: comparison with the Hawksley random zero sphygmomanometer under field conditions
- PMID: 1583434
- PMCID: PMC1059528
- DOI: 10.1136/jech.46.2.164
The Dinamap 1846SX automated blood pressure recorder: comparison with the Hawksley random zero sphygmomanometer under field conditions
Abstract
Study objective: The aim was to compare the performance of the Dinamap 1846SX automated oscillometric blood pressure recorder with that of the Hawksley random zero sphygmomanometer during use under field study conditions.
Design: Two independent within subject measurement comparisons were made, one in adults and one in children, each conducted in three stages over several months while the Dinamap instruments were being used in epidemiological field surveys.
Setting: The studies were done in outpatients clinics (adults) and primary schools (children).
Participants: 141 adults (20-85 years) and 152 children (5-7 years) took part.
Measurements and main results: In adults a pair of measurements was made with each instrument, the order alternating for consecutive subjects. In children one measurements was made with each instrument, in random order. Measurements with the Dinamap 1846SX were higher than those with the random zero sphygmomanometer both in adults (mean difference 8.1 mm Hg; 95% CI 6.5 to 9.7 mm Hg) and in children (mean difference 8.3 mm Hg; 95% CI 6.9 to 9.7 mm Hg). Diastolic measurements were on average very similar both in adults and in children. The results were consistent at all three stages of both studies. The differences in systolic measurement were independent of blood pressure level. However, the extent of agreement in diastolic pressure depended on the diastolic blood pressure level; in both studies Dinamap diastolic measurements were higher at low diastolic pressures while random zero diastolic measurements were higher at high diastolic pressures.
Conclusions: Systolic measurements made with the Dinamap 1846SX instrument are not directly comparable with those of the Hawksley random zero sphygmomanometer and are unlikely to be comparable with those of earlier Dinamap models. These differences have important implications for clinical practice and for comparisons of blood pressure measurement between epidemiological studies. However, the consistency of measurement by the Dinamap 1846SX over time suggests that the instrument may have a place in standardised blood pressure measurement in the research setting.
Similar articles
-
Random zero sphygmomanometer versus automatic oscillometric blood pressure monitor; is either the instrument of choice?J Hum Hypertens. 1995 Nov;9(11):885-9. J Hum Hypertens. 1995. PMID: 8583467 Clinical Trial.
-
Does the Hawksley random zero sphygmomanometer underestimate blood pressure, and by how much?J Hum Hypertens. 1995 May;9(5):337-43. J Hum Hypertens. 1995. PMID: 7623370 Clinical Trial.
-
Inaccuracy of the Hawksley random zero sphygmomanometer.Lancet. 1990 Dec 15;336(8729):1465-8. doi: 10.1016/0140-6736(90)93177-q. Lancet. 1990. PMID: 1979092 Clinical Trial.
-
Clinical evaluation of Dinamap 845 automated blood pressure recorder.Br Heart J. 1980 Feb;43(2):202-5. doi: 10.1136/hrt.43.2.202. Br Heart J. 1980. PMID: 7362713 Free PMC article.
-
Blood pressure measurement in childhood epidemiological studies.Circulation. 1995 Aug 15;92(4):1049-57. doi: 10.1161/01.cir.92.4.1049. Circulation. 1995. PMID: 7641339 Review.
Cited by
-
Size at birth, infant growth, and blood pressure at three years of age.J Pediatr. 2007 Dec;151(6):670-4. doi: 10.1016/j.jpeds.2007.05.010. Epub 2007 Aug 10. J Pediatr. 2007. PMID: 18035150 Free PMC article.
-
Assessing the impact of medication use on trends in major coronary risk factors in older British men: a cohort study.Eur J Cardiovasc Prev Rehabil. 2010 Oct;17(5):502-8. doi: 10.1097/HJR.0b013e3283378865. Eur J Cardiovasc Prev Rehabil. 2010. PMID: 20386311 Free PMC article.
-
Low prevalence of lipid lowering drug use in older men with established coronary heart disease.Heart. 2002 Jul;88(1):25-9. doi: 10.1136/heart.88.1.25. Heart. 2002. PMID: 12067936 Free PMC article.
-
Insulin resistance, haemostatic and inflammatory markers and coronary heart disease risk factors in Type 2 diabetic men with and without coronary heart disease.Diabetologia. 2004 Sep;47(9):1557-65. doi: 10.1007/s00125-004-1491-7. Epub 2004 Sep 9. Diabetologia. 2004. PMID: 15365613
-
Sex differences in body fat distribution and carotid intima media thickness: cross sectional survey using data from the British regional heart study.J Epidemiol Community Health. 2004 Aug;58(8):700-4. doi: 10.1136/jech.2003.014001. J Epidemiol Community Health. 2004. PMID: 15252075 Free PMC article.
References
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources