Magnetic resonance imaging of the upper abdomen using a free-breathing T2-weighted turbo spin echo sequence with navigator triggered prospective acquisition correction
- PMID: 15834908
- DOI: 10.1002/jmri.20293
Magnetic resonance imaging of the upper abdomen using a free-breathing T2-weighted turbo spin echo sequence with navigator triggered prospective acquisition correction
Abstract
Purpose: To evaluate a free-breathing navigator triggered T2-weighted turbo spin-echo sequence with prospective acquisition correction (T2w-PACE-TSE) for MRI of the upper abdomen in comparison to a conventional T2-weighted TSE (T2w-CTSE), a single-shot TSE (T2w-HASTE), and a T1-weighted gradient-echo sequence (T1w-FLASH).
Materials and methods: A total of 40 consecutive patients were examined at 1.5 T using free-breathing T2w-PACE-TSE, free-breathing T2w-CTSE, and breath-hold T2w-HASTE and T1w-FLASH acquisition. Images were evaluated qualitatively by three radiologists regarding motion artifacts, liver-spleen contrast, depiction of intrahepatic vessels, the pancreas and the adrenal glands, and overall image quality on a four-point scale. Quantitative analysis of the liver-spleen contrast was performed.
Results: Depiction and sharpness of intrahepatic vessels were rated significantly better (P < 0.01) using T2w-PACE-TSE compared to T2w-CTSE and T2w-HASTE sequences. Significantly higher contrast values were measured for T2w-PACE-TSE images compared to T2w-CTSE, T2w-HASTE, and T1w-FLASH images (P < 0.01). Mean examination time of the T2w-PACE-TSE was 7.91 minutes, acquisition time of the T2w-CTSE sequence was 4.52 minutes.
Conclusion: Prospective acquisition correction is an efficient method for reducing respiratory movement artifacts in T2w-TSE imaging of the upper abdomen. Compared to T2w-CTSE and T2w-HASTE sequences recognition of anatomical details and contrast can be significantly improved.
Similar articles
-
T2-weighted MRI of the upper abdomen: comparison of four fat-suppressed T2-weighted sequences including PROPELLER (BLADE) technique.Acad Radiol. 2010 Mar;17(3):368-74. doi: 10.1016/j.acra.2009.10.015. Epub 2009 Dec 30. Acad Radiol. 2010. PMID: 20042352
-
Magnetic resonance cholangiopancreatography using a free-breathing T2-weighted turbo spin-echo sequence with navigator-triggered prospective acquisition correction.Magn Reson Imaging. 2005 Nov;23(9):939-45. doi: 10.1016/j.mri.2005.07.002. Epub 2005 Oct 20. Magn Reson Imaging. 2005. PMID: 16310109
-
Image quality and focal lesion detection on T2-weighted MR imaging of the liver: comparison of two high-resolution free-breathing imaging techniques with two breath-hold imaging techniques.J Magn Reson Imaging. 2007 Aug;26(2):323-30. doi: 10.1002/jmri.21002. J Magn Reson Imaging. 2007. PMID: 17610287
-
Motion artifact control in body MR imaging.Magn Reson Imaging Clin N Am. 1999 May;7(2):289-301. Magn Reson Imaging Clin N Am. 1999. PMID: 10382162 Review.
-
[PSIF/CE-FAST (fast gradient echo) and TSE (turbo spin echo)].Nihon Rinsho. 1998 Nov;56(11):2773-7. Nihon Rinsho. 1998. PMID: 9847597 Review. Japanese.
Cited by
-
Liver metabolite concentrations measured with 1H MR spectroscopy.Radiology. 2012 Nov;265(2):565-75. doi: 10.1148/radiol.12112344. Epub 2012 Aug 13. Radiology. 2012. PMID: 22891360 Free PMC article.
-
MRI of the liver--how to do it.Pediatr Radiol. 2010 Apr;40(4):431-7. doi: 10.1007/s00247-009-1522-0. Pediatr Radiol. 2010. PMID: 20225098 Review.
-
With 3 Types of Respiratory Acquisition: 3.0 T Respiratory Triggered Acquisition Can Obtain Higher Quality DWI Images of the Upper Abdomen.Contrast Media Mol Imaging. 2022 Jul 9;2022:9579145. doi: 10.1155/2022/9579145. eCollection 2022. Contrast Media Mol Imaging. 2022. PMID: 35854769 Free PMC article.
-
Advantages of BioMatrix respiratory gating in free-breathing three-dimensional magnetic resonance cholangiopancreatography: a prospective comparative study.Insights Imaging. 2025 Jun 27;16(1):137. doi: 10.1186/s13244-025-02023-4. Insights Imaging. 2025. PMID: 40579676 Free PMC article.
-
Accelerated T2-weighted MRI of the liver at 3 T using a single-shot technique with deep learning-based image reconstruction: impact on the image quality and lesion detection.Abdom Radiol (NY). 2023 Jan;48(1):282-290. doi: 10.1007/s00261-022-03687-y. Epub 2022 Sep 28. Abdom Radiol (NY). 2023. PMID: 36171342 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical