Comparative analysis between epidural (Gaeltec) and subdural (Camino) intracranial pressure probes
- PMID: 1583546
- DOI: 10.1007/BF01617429
Comparative analysis between epidural (Gaeltec) and subdural (Camino) intracranial pressure probes
Abstract
The new fiberoptic Camino system has recently been introduced for clinical intracranial pressure (ICP) monitoring. We compared the subdural Camino system with the well-established epidural Gaeltec system in both in vitro and clinical conditions. In the in vitro experiments the intracranial vault was simulated by a tightly closed, fluid-filled box (0.9% sodium chloride) with the two probes inside. We simulated pulsating waveforms with a jet ventilator. No difference between the simulated curve patterns and values could be detected. In the clinical studies, both probes were implanted in 10 patients who had either head injuries, subarachnoid hemorrhage, or intracerebral hemorrhage. The in vivo comparison revealed no significant difference between the two systems in the recorded pressures in group 1 (ICP less than 20 mm Hg). The subdurally placed Camino probe showed insignificantly lower ICP values than did the extradural Gaeltec probe. Although group 2 (ICP greater than 20 mm Hg) waveforms were nearly identical, significant differences (p less than 0.01) in pressure measurements (systolic, diastolic, and mean) occurred (Camino, 18 +/- 3 mm Hg; Gaeltec, 27 +/- 3 mm Hg). Correlation coefficients for mean ICP values were 0.82 in group 1 and 0.49 in group 2. Problems with the Camino probe were usually mechanical and occurred in 2 patients. The problems were either easy to recognize or manifested as an ostensibly pathologic curve. No infection occurred with either system during or following implantation. The dissimilar characteristics of the two probes can be ascribed to their different extradural and subdural implantation sites. The Gaeltec probe was more durable over the period of implantation, which averaged 98 hours and ranged from 44 to 298 hours.(ABSTRACT TRUNCATED AT 250 WORDS)
Similar articles
-
Clinical evaluation of the Codman microsensor intracranial pressure monitoring system.Acta Neurochir Suppl. 1998;71:44-6. doi: 10.1007/978-3-7091-6475-4_14. Acta Neurochir Suppl. 1998. PMID: 9779140
-
The Camino intracranial pressure device in clinical practice: reliability, handling characteristics and complications.Acta Neurochir (Wien). 1998;140(11):1113-9; discussion 1119-20. doi: 10.1007/s007010050224. Acta Neurochir (Wien). 1998. PMID: 9870055
-
A comparison of extradural and intraparenchymatous intracranial pressures in head injured patients.Intensive Care Med. 1995 Oct;21(10):850-2. doi: 10.1007/BF01700971. Intensive Care Med. 1995. PMID: 8557876 Clinical Trial.
-
Intracranial pressure monitoring techniques.New Horiz. 1995 Aug;3(3):395-9. New Horiz. 1995. PMID: 7496747 Review.
-
The Brain Trauma Foundation. The American Association of Neurological Surgeons. The Joint Section on Neurotrauma and Critical Care. Recommendations for intracranial pressure monitoring technology.J Neurotrauma. 2000 Jun-Jul;17(6-7):497-506. doi: 10.1089/neu.2000.17.497. J Neurotrauma. 2000. PMID: 10937892 Review.
Cited by
-
Intracranial Pressure Monitoring: Invasive versus Non-Invasive Methods-A Review.Crit Care Res Pract. 2012;2012:950393. doi: 10.1155/2012/950393. Epub 2012 Jun 8. Crit Care Res Pract. 2012. PMID: 22720148 Free PMC article.
-
Evaluation of a fiberoptic system for airway pressure monitoring.J Clin Monit. 1994 Jul;10(4):247-50. doi: 10.1007/BF02899510. J Clin Monit. 1994. PMID: 7931455
-
Monitoring and Measurement of Intracranial Pressure in Pediatric Head Trauma.Front Neurol. 2020 Jan 14;10:1376. doi: 10.3389/fneur.2019.01376. eCollection 2019. Front Neurol. 2020. PMID: 32010042 Free PMC article. Review.
-
Advances in Intracranial Pressure Monitoring and Its Significance in Managing Traumatic Brain Injury.Int J Mol Sci. 2015 Dec 4;16(12):28979-97. doi: 10.3390/ijms161226146. Int J Mol Sci. 2015. PMID: 26690122 Free PMC article. Review.
-
Complications of brain tissue pressure monitoring with a fiberoptic device.Neurosurg Rev. 1998;21(4):254-9. doi: 10.1007/BF01105781. Neurosurg Rev. 1998. PMID: 10068186