Grids or air gaps for scatter reduction in digital radiography: a model calculation
Abstract
The relative advantages of grids and air gaps for scatter reduction in a digital radiography system were investigated using a theoretical model. In this model the properties of the scatter reduction device are described by primary transmission and selectivity. The signal-to-noise (SNR) improvement factor for fixed exposure to the patient was used as a performance indicator. The results show that the SNR improvement depends strongly on the local scatter fraction; for all practical configurations, however, it stays below a factor of 2. For high scatter fractions, an air gap of 20 cm has about the same effect on SNR improvement as a highly selective grid; for low and medium scatter conditions the air gap performs better than any grid. Additive system noise reduces the SNR improvement factor compared to the case with quantum noise only, the reduction being more pronounced for the grids than for the air gap. The results suggest that the use of an air gap instead of a grid is advantageous in digital radiography systems.
Similar articles
-
Studies of performance of antiscatter grids in digital radiography: effect on signal-to-noise ratio.Med Phys. 1990 Jul-Aug;17(4):655-64. doi: 10.1118/1.596496. Med Phys. 1990. PMID: 2215411
-
Experimental evaluation of fiber-interspaced antiscatter grids for large patient imaging with digital x-ray systems.Phys Med Biol. 2007 Aug 21;52(16):4863-80. doi: 10.1088/0031-9155/52/16/010. Epub 2007 Jul 30. Phys Med Biol. 2007. PMID: 17671340
-
Physical evaluation of prototype high-performance anti-scatter grids: potential for improved digital radiographic image quality.Phys Med Biol. 2009 Jan 21;54(2):N37-42. doi: 10.1088/0031-9155/54/2/N02. Epub 2008 Dec 19. Phys Med Biol. 2009. PMID: 19098352
-
Electronic scanning-slit fluorography.Acta Radiol Suppl. 1989;373:1-48. Acta Radiol Suppl. 1989. PMID: 2488063 Review.
-
Contrast and scatter in x-ray imaging.Radiographics. 1991 Mar;11(2):307-23. doi: 10.1148/radiographics.11.2.2028065. Radiographics. 1991. PMID: 2028065 Review.
Cited by
-
Monte Carlo evaluation of scatter mitigation strategies in cone-beam CT.Med Phys. 2010 Oct;37(10):5456-70. doi: 10.1118/1.3488978. Med Phys. 2010. PMID: 21089782 Free PMC article.
-
A simulation study to evaluate the effect of 2D antiscatter grid primary transmission on flat panel detector based CBCT image quality.Biomed Phys Eng Express. 2023 Oct 4;9(6):10.1088/2057-1976/acfb8a. doi: 10.1088/2057-1976/acfb8a. Biomed Phys Eng Express. 2023. PMID: 37729884 Free PMC article.
-
A new software scheme for scatter correction based on a simple radiographic scattering model.Med Biol Eng Comput. 2019 Feb;57(2):489-503. doi: 10.1007/s11517-018-1893-1. Epub 2018 Sep 19. Med Biol Eng Comput. 2019. PMID: 30232700
-
Antiscatter grid use in pediatric digital tomosynthesis imaging.J Appl Clin Med Phys. 2011 Nov 15;12(4):3641. doi: 10.1120/jacmp.v12i4.3641. J Appl Clin Med Phys. 2011. PMID: 22089021 Free PMC article.
-
Intraoperative cone-beam and slot-beam CT: 3D image quality and dose with a slot collimator on the O-arm imaging system.Med Phys. 2021 Nov;48(11):6800-6809. doi: 10.1002/mp.15221. Epub 2021 Sep 23. Med Phys. 2021. PMID: 34519364 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous