Escherichia coli O157:H7 detection limit of millimeter-sized PZT cantilever sensors is 700 cells/mL
- PMID: 15844324
- DOI: 10.2116/analsci.21.355
Escherichia coli O157:H7 detection limit of millimeter-sized PZT cantilever sensors is 700 cells/mL
Abstract
A composite self-excited millimeter-sized lead zirconate titanate (PZT) glass cantilever (2 mm x 1.8 mm; sensing area of 6 mm2) was fabricated for the detection of Escherichia coli (E. coli) O157:H7. The fundamental and second mode resonance in air was 10.95 +/- 0.05 kHz and 43.45 +/- 0.05 kHz, respectively. Affinity purified monoclonal antibody (anti-E. coli O157:H7) specific to the pathogen E. coli O157:H7 was immobilized at the cantilever glass tip, and then immersed in liquid containing the pathogen (70 to 7 x 10(7) cells/mL). The resonant frequency showed a reduction and reached a steady state shift of 0 +/- 5, 46 +/- 5, 260 +/- 5, and 1010 +/- 5 Hz corresponding to 0, 700, 7000, and 7 x 10(7) cells/mL. From the experiments conducted, the detection limit of the sensor was 700 cells/mL.
Similar articles
-
Detection of pathogen Escherichia coli O157:H7 using self-excited PZT-glass microcantilevers.Biosens Bioelectron. 2005 Sep 15;21(3):462-73. doi: 10.1016/j.bios.2004.11.009. Epub 2004 Dec 21. Biosens Bioelectron. 2005. PMID: 16076436
-
Detect of Escherichia coli O157:H7 in ground beef samples using piezoelectric excited millimeter-sized cantilever (PEMC) sensors.Biosens Bioelectron. 2007 Feb 15;22(7):1296-302. doi: 10.1016/j.bios.2006.05.028. Epub 2006 Jul 10. Biosens Bioelectron. 2007. PMID: 16831542
-
An anti E. coli O157:H7 antibody-immobilized microcantilever for the detection of Escherichia coli (E. coli).Anal Sci. 2004 Apr;20(4):585-7. doi: 10.2116/analsci.20.585. Anal Sci. 2004. PMID: 15116951
-
Detection and quantification of proteins using self-excited PZT-glass millimeter-sized cantilever.Biosens Bioelectron. 2005 Oct 15;21(4):597-607. doi: 10.1016/j.bios.2004.12.016. Epub 2005 Jan 22. Biosens Bioelectron. 2005. PMID: 16202873
-
Development of a low-cost paper-based ELISA method for rapid Escherichia coli O157:H7 detection.Anal Biochem. 2018 Feb 1;542:58-62. doi: 10.1016/j.ab.2017.11.010. Epub 2017 Nov 20. Anal Biochem. 2018. PMID: 29158131 Review.
Cited by
-
Precise mass determination of single cell with cantilever-based microbiosensor system.PLoS One. 2017 Nov 21;12(11):e0188388. doi: 10.1371/journal.pone.0188388. eCollection 2017. PLoS One. 2017. PMID: 29161333 Free PMC article.
-
Development of a Microforce Sensor and Its Array Platform for Robotic Cell Microinjection Force Measurement.Sensors (Basel). 2016 Apr 6;16(4):483. doi: 10.3390/s16040483. Sensors (Basel). 2016. PMID: 27058545 Free PMC article.
-
Sensor to detect endothelialization on an active coronary stent.Biomed Eng Online. 2010 Nov 4;9:67. doi: 10.1186/1475-925X-9-67. Biomed Eng Online. 2010. PMID: 21050471 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources