Assessment of coronary artery stents by 16 slice computed tomography
- PMID: 15845613
- PMCID: PMC1860967
- DOI: 10.1136/hrt.2004.056010
Assessment of coronary artery stents by 16 slice computed tomography
Abstract
Objective: To analyse coronary stents with multislice spiral computed tomography (MSCT) in comparison with coronary angiography.
Patients and methods: 310 patients referred for conventional coronary angiography underwent MSCT on the next day (16 x 0.75 mm cross section, 420 ms rotation, 110 ml contrast agent intravenously at 4 ml/s). Two independent blinded reviewers analysed the MSCT.
Results: 143 patients had previous stenting (232 stents) and 190 (82%) of the 232 stents were detected. Intrastent lumen was interpretable in 126 (64%) of the detected stents. Lumen interpretability depended on stent diameter: for stent diameter > 3 mm, 81% of lumens were interpretable, as against 51% with < or = 3 mm stent diameter (p < 0.001). Restenosis detection likewise depended on stent diameter: with small stents (< or = 3 mm), sensitivity and specificity of MSCT were 54% and 100%, respectively; positive and negative predictive values were 100% and 94%. For stents with > 3 mm diameter, corresponding values were 86%, 100%, 100%, and 99%.
Conclusion: 16 slice MSCT allows analysis of in-stent lumen in about half of all stented angioplasties. It performs better when stent diameter is more than 3 mm and may offer a non-invasive alternative to conventional coronary angiography for monitoring stented coronary arteries. Technical progress may improve interpretability and hence increase the yield of MSCT in this application.
References
-
- Zellweger M J, Weinbacher M, Zutter A W.et al Long‐term outcome of patients with silent versus symptomatic ischemia six months after percutaneous coronary intervention and stenting. J Am Coll Cardiol 20034233–40. - PubMed
-
- Nieman K, Cademartiri F, Lemos P A.et al Reliable noninvasive coronary angiography with fast submillimeter multislice spiral computed tomography. Circulation 20021062051–2054. - PubMed
-
- Ropers D, Baum U, Pohle K.et al Detection of coronary artery stenoses with thin‐slice multi‐detector row spiral computed tomography and multiplanar reconstruction. Circulation 2003107664–666. - PubMed
-
- Nieaman K, Cademartiri F, Raaijmakers R.et al Noninvasive angiography evaluation of coronary stents with multi‐slice spiral computed tomography. Herz 200328136–142. - PubMed
-
- Schuijf J D, Bax J J, Jukema W.et al Feasibility of assessment of coronary stent patency using 16‐slice computed tomography. Am J Cardiol 200494427–430. - PubMed