Patient, procedural, and hardware factors associated with pacemaker lead failures in pediatrics and congenital heart disease
- PMID: 15851146
- DOI: 10.1016/j.hrthm.2004.02.020
Patient, procedural, and hardware factors associated with pacemaker lead failures in pediatrics and congenital heart disease
Abstract
Objectives: To examine outcomes of children with pacemakers over a 22-year period and identify risk factors for lead failure.
Background: Small patient size, structural cardiac abnormalities, and growth may complicate pediatric pacemaker management. Better knowledge of risk factors for lead failure in these patients may help improve future outcomes.
Methods: All pacemaker patients followed at one pediatric center 1980-2002 were included. Lead failures were identified retrospectively as leads repaired, replaced, or abandoned due to fracture, insulation break, dislodgement, or abnormalities in pacing or sensing. Risk factors were identified using logistic regression and Cox analyses.
Results: A total of 1007 leads were implanted in 497 patients during the study period (5175 lead-years). Median age at implant was 9 years (0-55); 64% of patients had structural congenital heart disease. Median follow-up time was 6.2 years (0-22). Lead failure occurred in 155 leads (15%), and 115 patients (23%), with 28% of patients experiencing multiple failures. Significant independent correlates of lead failure included age <12 years at implant, history of structural congenital heart defects, and epicardial lead placement. Younger patients (<12 years) experienced significantly more lead fractures than older children (P = .005), while patients with congenital heart defects experienced more exit block. Epicardial leads were more likely to fail due to fracture or exit block, while transvenous leads failed more due to insulation breaks or dislodgements.
Conclusions: Pediatric pacing patients have a high incidence lead failures. These occur most commonly in younger patients, structural congenital heart disease, and those with epicardial lead systems. Approaches to pacing system implantation and follow-up in these patients need to be individualized, with special attention to minimizing risk of lead failures. Our findings suggest that expanded utilization of transvenous systems in smaller patients seems justified when anatomy permits.
Similar articles
-
Predictors affecting durability of epicardial pacemaker leads in pediatric patients.J Thorac Cardiovasc Surg. 2008 Feb;135(2):361-6. doi: 10.1016/j.jtcvs.2007.09.002. J Thorac Cardiovasc Surg. 2008. PMID: 18242269
-
Efficacy of prophylactic epicardial pacing leads in children and young adults.Ann Thorac Surg. 2004 Jul;78(1):197-202; discussion 202-3. doi: 10.1016/j.athoracsur.2004.02.008. Ann Thorac Surg. 2004. PMID: 15223427 Review.
-
Pacing activity, patient and lead survival over 20 years of permanent epicardial pacing in children.Ann Thorac Surg. 2004 Apr;77(4):1366-70. doi: 10.1016/j.athoracsur.2003.08.057. Ann Thorac Surg. 2004. PMID: 15063268
-
Long-term follow-up shows excellent transmural atrial lead performance in patients with complex congenital heart disease.Circ Arrhythm Electrophysiol. 2014 Aug;7(4):652-7. doi: 10.1161/CIRCEP.113.001321. Epub 2014 Jun 6. Circ Arrhythm Electrophysiol. 2014. PMID: 24907290
-
Technical aspects of pacing in adult and pediatric congenital heart disease.Pacing Clin Electrophysiol. 2008 Feb;31 Suppl 1:S28-31. doi: 10.1111/j.1540-8159.2008.00951.x. Pacing Clin Electrophysiol. 2008. PMID: 18226031 Review.
Cited by
-
Pacing device therapy in infants and children: a review.J Artif Organs. 2013 Mar;16(1):23-33. doi: 10.1007/s10047-012-0668-y. Epub 2012 Oct 27. J Artif Organs. 2013. PMID: 23104398 Review.
-
Harnessing cell reprogramming for cardiac biological pacing.J Biomed Sci. 2023 Aug 26;30(1):74. doi: 10.1186/s12929-023-00970-y. J Biomed Sci. 2023. PMID: 37633890 Free PMC article. Review.
-
A case of pacemaker dysfunction due to interference from a stent placed in the subclavian vein.HeartRhythm Case Rep. 2021 Jul 30;7(11):722-725. doi: 10.1016/j.hrcr.2021.07.014. eCollection 2021 Nov. HeartRhythm Case Rep. 2021. PMID: 34820266 Free PMC article. No abstract available.
-
Right-Sided Subcutaneous Implantable Cardioverter Defibrillator System Implantation in a Patient with Complex Congenital Heart Disease and Dextrocardia: A Case Report and Literature Review.Case Rep Cardiol. 2019 Feb 5;2019:3907190. doi: 10.1155/2019/3907190. eCollection 2019. Case Rep Cardiol. 2019. PMID: 30867968 Free PMC article.
-
Temporary transvenous atrioventricular synchronous pacing using a single lead in a pediatric patient.HeartRhythm Case Rep. 2019 Oct 1;5(12):593-596. doi: 10.1016/j.hrcr.2019.09.008. eCollection 2019 Dec. HeartRhythm Case Rep. 2019. PMID: 31890584 Free PMC article. No abstract available.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical