Individually adapted examination protocols for reduction of radiation exposure for 16-MDCT chest examinations
- PMID: 15855093
- DOI: 10.2214/ajr.184.5.01841437
Individually adapted examination protocols for reduction of radiation exposure for 16-MDCT chest examinations
Abstract
Objective: The purpose of our study was to develop a simple protocol for reduction of radiation exposure without loss of diagnostic information in chest 16-MDCT.
Materials and methods: Two hundred and four patients underwent MDCT of the thorax (Somatom Sensation 16, Siemens). Group 1 was scanned using a standard protocol with 100 mAs(effective) (mAs(eff)). Group 2 was scanned using a dose modulation template (CareDose). Group 3 was scanned with mAs(eff) = body weight (kg). Group 4 was scanned with a combination of weight-adapted mAs(eff) and dose modulation. All other parameters were kept constant. Signal-to-noise ratio was assessed as an objective measurement for image quality, and subjective image quality was rated by three experienced radiologists on a 4-point scale. Effective dose was calculated using dedicated software.
Results: The mean noise measurement values were 8.31 H for the 100 mAs(eff) protocol for the regression between weight and signal-to-noise (p < 0.0001), 9.08 H for group 2 (p < 0.0001), 9.0 H for group 3 (p = 0.5051), and 9.98 H for group 4 (p = 0.0152). The median image quality was 1 (1 = highest quality) in all subgroups. The mean effective dose was 6.83 mSv, 5.92 mSv, 4.73 mSv, and 3.97 mSv, respectively. The least correlation between weight and image noise was achieved for the individually weight-adapted protocol and in the weight-adapted with CareDose combination.
Conclusion: By tube current time product adaptation (kg = mAs(eff)) combined with an online tube current modulation template, a well-balanced examination without significant loss of information was achieved for this specific scanner. Thus, individually adapted protocols for chest 16-MDCT can be recommended.
Similar articles
-
Individually adapted examination protocols for reduction of radiation exposure in chest CT.Invest Radiol. 2001 Oct;36(10):604-11. doi: 10.1097/00004424-200110000-00006. Invest Radiol. 2001. PMID: 11577271
-
Multidetector computed tomography chest examinations with low-kilovoltage protocols in adults: effect on image quality and radiation dose.J Comput Assist Tomogr. 2009 May-Jun;33(3):416-21. doi: 10.1097/RCT.0b013e318181fab5. J Comput Assist Tomogr. 2009. PMID: 19478637 Clinical Trial.
-
Radiation dose reduction by using 100-kV tube voltage in cardiac 64-slice computed tomography: a comparative study.Eur J Radiol. 2010 Jul;75(1):e51-6. doi: 10.1016/j.ejrad.2009.07.012. Epub 2009 Aug 9. Eur J Radiol. 2010. PMID: 19671491
-
Saving dose in triple-rule-out computed tomography examination using a high-pitch dual spiral technique.Invest Radiol. 2010 Feb;45(2):64-71. doi: 10.1097/RLI.0b013e3181c15842. Invest Radiol. 2010. PMID: 20027121
-
Automated computed tomography dose-saving algorithm to protect radiosensitive tissues: estimation of radiation exposure and image quality considerations.Invest Radiol. 2012 Feb;47(2):148-52. doi: 10.1097/RLI.0b013e3182311504. Invest Radiol. 2012. PMID: 21934513
Cited by
-
Tube current reduction in pediatric non-ECG-gated heart CT by combined tube current modulation.Pediatr Radiol. 2006 Apr;36(4):344-51. doi: 10.1007/s00247-005-0105-y. Epub 2006 Feb 25. Pediatr Radiol. 2006. PMID: 16501970 Clinical Trial.
-
Radiation dose estimates in dual-source computed tomography coronary angiography.Eur Radiol. 2008 Mar;18(3):592-9. doi: 10.1007/s00330-007-0786-8. Epub 2007 Oct 2. Eur Radiol. 2008. PMID: 17909816
-
Evaluation of automated attenuation-based tube current adaptation for coronary calcium scoring in MDCT in a cohort of 262 patients.Eur Radiol. 2007 Jul;17(7):1850-7. doi: 10.1007/s00330-006-0543-4. Epub 2007 Feb 17. Eur Radiol. 2007. PMID: 17308926
-
Low-dose ECG-gated 64-slices helical CT angiography of the chest: evaluation of image quality in 105 patients.Eur Radiol. 2006 Oct;16(10):2137-46. doi: 10.1007/s00330-006-0218-1. Epub 2006 Apr 12. Eur Radiol. 2006. PMID: 16609862
-
Achieving consistent image quality with dose optimization in 64-row multidetector computed tomography prospective ECG gated coronary calcium scoring.Int J Cardiovasc Imaging. 2011 Apr;27(4):611-7. doi: 10.1007/s10554-010-9701-5. Epub 2010 Sep 18. Int J Cardiovasc Imaging. 2011. PMID: 20852938
MeSH terms