Clinical Trial
. 2005 Jul;25(7):453-8.
doi: 10.1038/sj.jp.7211325.

Work of breathing during nasal continuous positive airway pressure in preterm infants: a comparison of bubble vs variable-flow devices

Affiliations
Clinical Trial

Work of breathing during nasal continuous positive airway pressure in preterm infants: a comparison of bubble vs variable-flow devices

Ellina Liptsen et al. J Perinatol. 2005 Jul.

Abstract

Objective: To compare work of breathing and breathing asynchrony during bubble nasal continuous positive airway pressure (NCPAP) vs variable-flow (VF)-NCPAP in premature infants.

Study design: We studied 18 premature infants of birth weight <1500 g who required NCPAP for mild respiratory distress. Each infant was studied on bubble and VF-NCPAP at 8, 6, 4, and 0 cm H2O. Tidal volumes were obtained by calibrated respiratory inductance plethysmography. Esophageal pressure estimated intrapleural pressure. Inspiratory and resistive work of breathing were calculated from pressure-volume data. Breathing asynchrony was assessed with phase angle. The results at all NCPAP levels were referenced to VF-NCPAP values at 8 cm H2O.

Results: Provision of NCPAP with either device decreased inspiratory work of breathing, tidal volume, and minute ventilation relative to NCPAP of 0 cm H2O. Bubble NCPAP did not decrease resistive work of breathing relative to 0 cm H2O. Resistive work of breathing (p=0.01), respiratory rate (p<0.03), and phase angle (p=0.002) were all greater with bubble compared to VF-NCPAP.

Conclusion: The more labored and asynchronous breathing seen with bubble NCPAP may lead to higher failure rates over the long term than with VF-NCPAP.