Evaluation of accuracy of pulse oximetry in dogs
Abstract
The accuracy of a pulse oximeter was evaluated over a wide range of arterial oxygen and carbon dioxide tensions, using 2 probes (finger probe and ear probe) and 2 monitoring sites (tongue and tail) in anesthetized dogs. The arterial oxygen saturation of hemoglobin (SaO2) measured directly with a multiwavelength spectrophotometer was compared with saturation estimated by pulse oximetry (SpO2). Linear regression analysis of the pooled data from 399 simultaneous measurements of SpO2 and SaO2 indicated a highly significant correlation of SpO2 with SaO2 (r = 0.97; P less than or equal to 0.0001). Although the mean difference (+/- SD) between SpO2 and SaO2 for pooled data was small (-0.06 +/- 6.8%), SpO2 tended to underestimate high SaO2 values (greater than or equal to 70%) and to overestimate low SaO2 values (less than 70%). When SaO2 values were greater than or equal to 70%, the ear probe applied to the tail was less accurate (produced a significantly greater SpO2-SaO2 difference) than the ear probe on the tongue, or the finger probe at either site. When SaO2 values were less than or equal to 50%, the finger probe applied at the tail was more accurate (produced significantly smaller SpO2-SaO2 differences) than the ear probe at either site. When SaO2 values were less than or equal to 70%, high arterial carbon dioxide tension (greater than or equal to 60 mm of Hg) was associated with greater overestimation of SaO2.
Similar articles
-
Evaluation of pulse oximeter probes and probe placement in healthy dogs.J Am Anim Hosp Assoc. 1995 Jan-Feb;31(1):9-14. doi: 10.5326/15473317-31-1-9. J Am Anim Hosp Assoc. 1995. PMID: 7820772
-
[Accuracy of a pulse oximeter during hypoxia].Masui. 1996 Apr;45(4):479-82. Masui. 1996. PMID: 8725606 Japanese.
-
Buccal pulse oximeter is more accurate than finger pulse oximeter in measuring oxygen saturation.Anesth Analg. 1992 Oct;75(4):495-8. Anesth Analg. 1992. PMID: 1530160
-
Pulse oximetry: fundamentals and technology update.Med Devices (Auckl). 2014 Jul 8;7:231-9. doi: 10.2147/MDER.S47319. eCollection 2014. Med Devices (Auckl). 2014. PMID: 25031547 Free PMC article. Review.
-
The accuracy of pulse oximetry in measuring oxygen saturation by levels of skin pigmentation: a systematic review and meta-analysis.BMC Med. 2022 Aug 16;20(1):267. doi: 10.1186/s12916-022-02452-8. BMC Med. 2022. PMID: 35971142 Free PMC article.
Cited by
-
The use of High-Flow Nasal Oxygen Therapy in 4 dogs undergoing bronchoscopy.Front Vet Sci. 2023 Mar 30;10:1088103. doi: 10.3389/fvets.2023.1088103. eCollection 2023. Front Vet Sci. 2023. PMID: 37065223 Free PMC article.
-
Anesthetic Effects of Intramuscular Alfaxalone-Ketamine in Naked Mole Rats (Heterocephalus glaber).J Am Assoc Lab Anim Sci. 2020 Sep 1;59(5):539-545. doi: 10.30802/AALAS-JAALAS-19-000170. Epub 2020 Jul 17. J Am Assoc Lab Anim Sci. 2020. PMID: 32680579 Free PMC article.
-
Assessment of clinical application of pulse oximetry probes in llamas and alpacas.Vet Med Sci. 2017 Jul 11;3(3):169-175. doi: 10.1002/vms3.68. eCollection 2017 Aug. Vet Med Sci. 2017. PMID: 29067213 Free PMC article.
-
The pharmacological effects of the anesthetic alfaxalone after intramuscular administration to dogs.J Vet Med Sci. 2015 Mar;77(3):289-96. doi: 10.1292/jvms.14-0368. Epub 2014 Nov 27. J Vet Med Sci. 2015. PMID: 25428797 Free PMC article.
-
Relationship between Respiratory Rate, Oxygen Saturation, and Blood Test Results in Dogs with Chronic or Acute Respiratory Disease: A Retrospective Study.Vet Sci. 2024 Jan 10;11(1):27. doi: 10.3390/vetsci11010027. Vet Sci. 2024. PMID: 38250933 Free PMC article.