Inspiratory muscle work of breathing during flow-by, demand-flow, and continuous-flow systems in patients with chronic obstructive pulmonary disease
- PMID: 1586070
- DOI: 10.1164/ajrccm/145.5.1219
Inspiratory muscle work of breathing during flow-by, demand-flow, and continuous-flow systems in patients with chronic obstructive pulmonary disease
Abstract
The effect of flow-by continuous positive airway pressure (CPAP) in comparison with continuous-flow (CF) CPAP on inspiratory muscle work of breathing (WI) in intubated patients is not known. We hypothesized that WI during flow-by CPAP would be comparable with that during CF CPAP. In nine intubated male patients recovering from acute respiratory failure related to chronic obstructive pulmonary disease, we compared the effects of flow-by, demand-flow, and CF CPAP on WI. We also evaluated the extent to which the addition of 5 cm H2O of pressure support to demand-flow CPAP (DF-PS5) decreases WI. At CPAP levels of zero and 8 cm H2O, flow-by, demand-flow, DF-PS5 were applied in random order followed by CF CPAP for 15 min each. WI (expressed as Joules per liter and Joules per minute), maximal airway pressure drop during inspiration (delta Paw), total lung resistance (RL), lung compliance (CL), and ventilatory variables were measured. At CPAP of zero cm H2O, WI with flow-by was comparable with CF CPAP, and significantly less than with demand-flow CPAP. At both CPAP of zero and 8 cm H2O, the addition of 5 cm H2O PS to demand-flow CPAP reduced WI significantly to a level comparable with that of flow-by CPAP. At both CPAP levels, delta Paw was the largest, with demand-flow CPAP. RL and CL were not significantly different between the different CPAP systems. We conclude that WI with flow-by CPAP is comparable with that with CF CPAP, and significantly less than with demand-flow CPAP.(ABSTRACT TRUNCATED AT 250 WORDS)
Similar articles
-
Characteristics of the ventilator pressure- and flow-trigger variables.Intensive Care Med. 1995 Feb;21(2):159-68. doi: 10.1007/BF01726540. Intensive Care Med. 1995. PMID: 7775698 Review.
-
Continuous positive airway pressure reduces work of breathing and dyspnea during weaning from mechanical ventilation in severe chronic obstructive pulmonary disease.Am Rev Respir Dis. 1990 Feb;141(2):281-9. doi: 10.1164/ajrccm/141.2.281. Am Rev Respir Dis. 1990. PMID: 2405757
-
Effect of low-level PEEP on inspiratory work of breathing in intubated patients, both with healthy lungs and with COPD.Intensive Care Med. 1995 Nov;21(11):887-95. doi: 10.1007/BF01712329. Intensive Care Med. 1995. PMID: 8636520 Clinical Trial.
-
Nasal continuous positive airway pressure facilitates respiratory muscle function during sleep in severe chronic obstructive pulmonary disease.Am Rev Respir Dis. 1991 May;143(5 Pt 1):928-35. doi: 10.1164/ajrccm/143.5_Pt_1.928. Am Rev Respir Dis. 1991. PMID: 2024845
-
Inspiratory flow rate: more may not be better.Crit Care Med. 1999 Apr;27(4):670-1. doi: 10.1097/00003246-199904000-00002. Crit Care Med. 1999. PMID: 10321642 Review. No abstract available.
Cited by
-
Consensus conference on mechanical ventilation--January 28-30, 1993 at Northbrook, Illinois, USA. Part I. European Society of Intensive Care Medicine, the ACCP and the SCCM.Intensive Care Med. 1994;20(1):64-79. doi: 10.1007/BF02425061. Intensive Care Med. 1994. PMID: 8163765 Review. No abstract available.
-
Inspiratory muscle training, with or without concomitant pulmonary rehabilitation, for chronic obstructive pulmonary disease (COPD).Cochrane Database Syst Rev. 2023 Jan 6;1(1):CD013778. doi: 10.1002/14651858.CD013778.pub2. Cochrane Database Syst Rev. 2023. PMID: 36606682 Free PMC article.
-
Superimposed inspiratory work of the Siemens Servo 300 ventilator during continuous positive airway pressure.Intensive Care Med. 1995 Dec;21(12):1023-6. doi: 10.1007/BF01700665. Intensive Care Med. 1995. PMID: 8750128 Clinical Trial.
-
Difficult weaning.Intensive Care Med. 1993;19 Suppl 2:S69-73. doi: 10.1007/BF01708804. Intensive Care Med. 1993. PMID: 8106681 Review. No abstract available.
-
Characteristics of the ventilator pressure- and flow-trigger variables.Intensive Care Med. 1995 Feb;21(2):159-68. doi: 10.1007/BF01726540. Intensive Care Med. 1995. PMID: 7775698 Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous