Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2005 Apr 29;37(5):1081-8.
doi: 10.1016/j.jpba.2004.07.041.

Development and validation of a bioanalytical method using automated solid-phase extraction and LC-UV for the simultaneous determination of lumefantrine and its desbutyl metabolite in plasma

Affiliations
Comparative Study

Development and validation of a bioanalytical method using automated solid-phase extraction and LC-UV for the simultaneous determination of lumefantrine and its desbutyl metabolite in plasma

N Lindegårdh et al. J Pharm Biomed Anal. .

Abstract

A bioanalytical method for the determination of lumefantrine (LF) and its metabolite desbutyl-lumefantrine (DLF) in plasma by solid-phase extraction (SPE) and liquid chromatography has been developed. Plasma proteins were precipitated with acetonitrile:acetic acid (99:1, v/v) containing a DLF analogue internal standard before being loaded onto a octylsilica (3 M Empore) SPE column. Two different DLF analogues were evaluated as internal standards. The compounds were analysed by liquid chromatography UV detection on a SB-CN (250 mm x 4.6 mm) column with a mobile phase containing acetonitrile-sodium phosphate buffer pH (2.0; 0.1 M) (55:45, v/v) and sodium perchlorate 0.05 M. Different SPE columns were evaluated during method development to optimise reproducibility and recovery for LF, DLF and the two different DLF analogues. The within-day precisions for LF were 6.6 and 2.1% at 0.042 and 8.02 microg/mL, respectively, and for DLF 4.5 and 1.5% at 0.039 and 0.777 microg/mL, respectively. The between-day precisions for LF were 12.0 and 2.9% at 0.042 and 8.02 microg/mL, respectively, while for DLF 0.7 and 1.2% at 0.039 and 0.777 microg/mL, respectively. The limit of quantification was 0.024 and 0.021 microg/mL for LF and DLF, respectively. Different amounts of lipids in plasma did not affect the absolute recovery of LF or DLF.

PubMed Disclaimer

Publication types

LinkOut - more resources