Optical coherence tomography of the Ex-PRESS miniature glaucoma implant
Abstract
Purpose: To describe the localisation of the Ex-PRESS miniature glaucoma implant with an experimental setup for optical coherence tomography (OCT) of the anterior segment of the eye.
Methods: An OCT scanner, central wavelength 1,280 nm, bandwidth 60 nm, resolution of 12 microm, was built onto a slitlamp to scan the anterior segment of the eye. Five ex-vivo porcine eyes received an Ex-PRESS miniature glaucoma implant and were used as a model to visualise the position of the implant in the anterior segment.
Results: In the ex-vivo porcine eyes, the OCT images showed the anatomy of the anterior segment in great detail. The anterior segment OCT was able to visualise the whole outline and position of the implant. The abrupt change in reflectivity going from tissue to the implant resulted in a clear border along the circumference of the whole device.
Conclusion: In this paper, we have shown that we were able to outline the Ex-PRESS miniature glaucoma implant in the anterior segment of the ex-vivo porcine eye by using an experimental OCT setup built onto a slitlamp. The acquisition time of 0.8 s is short enough to allow for the scanning of patients, and anterior segment OCT is expected to aid in providing answers to the question regarding which parameters will determine the success or failure of such a device.
Similar articles
-
Visualization of blebs using anterior-segment optical coherence tomography after glaucoma drainage implant surgery.Ophthalmology. 2013 May;120(5):978-83. doi: 10.1016/j.ophtha.2012.10.015. Epub 2013 Jan 21. Ophthalmology. 2013. PMID: 23347982
-
EX-PRESS® Implant Position and Function: Comparative Evaluation with Ultrasound Biomicroscopy and Optical Coherence Tomography.Middle East Afr J Ophthalmol. 2016 Jan-Mar;23(1):110-4. doi: 10.4103/0974-9233.171774. Middle East Afr J Ophthalmol. 2016. PMID: 26957849 Free PMC article.
-
Application of anterior segment optical coherence tomography in glaucoma.Surv Ophthalmol. 2014 May-Jun;59(3):311-27. doi: 10.1016/j.survophthal.2013.06.005. Epub 2013 Oct 15. Surv Ophthalmol. 2014. PMID: 24138894 Review.
-
The use of anterior segment optical coherence tomography in glaucoma drainage implant surgery.Ophthalmic Surg Lasers Imaging. 2008 Mar-Apr;39(2):164-6. doi: 10.3928/15428877-20080301-01. Ophthalmic Surg Lasers Imaging. 2008. PMID: 18435346
-
[Anterior segment optical coherence tomography in glaucoma].Klin Monbl Augenheilkd. 2008 Mar;225(3):194-9. doi: 10.1055/s-2008-1027241. Klin Monbl Augenheilkd. 2008. PMID: 18351532 Review. German.
Cited by
-
Anterior segment optical coherence tomography imaging of conjunctival filtering blebs after glaucoma surgery.Biomed Res Int. 2014;2014:610623. doi: 10.1155/2014/610623. Epub 2014 Jul 20. Biomed Res Int. 2014. PMID: 25136603 Free PMC article. Review.
-
Imaging of filtering blebs after implantation of the Ex-PRESS shunt with the use of the Visante optical coherence tomography.Int J Ophthalmol. 2015 Jun 18;8(3):492-5. doi: 10.3980/j.issn.2222-3959.2015.03.10. eCollection 2015. Int J Ophthalmol. 2015. PMID: 26085996 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical