Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Clinical Trial
. 2005 May;129(5):1056-63.
doi: 10.1016/j.jtcvs.2004.08.028.

Exercise hemodynamics of bovine versus porcine bioprostheses: a prospective randomized comparison of the mosaic and perimount aortic valves

Affiliations
Free article
Clinical Trial

Exercise hemodynamics of bovine versus porcine bioprostheses: a prospective randomized comparison of the mosaic and perimount aortic valves

Walter B Eichinger et al. J Thorac Cardiovasc Surg. 2005 May.
Free article

Abstract

Objective: This prospective randomized study compares a porcine with a bovine bioprosthesis in the aortic position with regard to hemodynamic performance during exercise.

Methods: Between August of 2000 and December of 2002, 136 patients underwent aortic valve replacement with the porcine Medtronic Mosaic (n = 66) or the bovine Carpentier-Edwards Perimount (n = 70) bioprosthesis. Transthoracic echocardiography was performed to assess hemodynamic and dimensional data preoperatively and 10 months postoperatively; the latter follow-up included stress echocardiography with treadmill exercise.

Results: At rest and during exercise (25 and 50 W), there was a significant difference in mean pressure gradient between the bovine and the porcine valves with labeled sizes 21 and 23, with superiority of the Perimount prosthesis. There was no difference in effective orifice area and incidence of patient-prosthesis mismatch among all sizes. The left ventricular mass index decreased significantly within 10 months postoperatively in the size 23 bovine group and the size 25 porcine group.

Conclusions: Our data show a significant superiority of pressure gradients for the bovine bioprosthesis, especially with small valve sizes, when compared with the porcine device, which is more distinctive during exercise.

PubMed Disclaimer

Comment in

  • Prosthetic aortic valve replacement.
    Sievers HH. Sievers HH. J Thorac Cardiovasc Surg. 2005 May;129(5):961-5. doi: 10.1016/j.jtcvs.2004.12.036. J Thorac Cardiovasc Surg. 2005. PMID: 15867766 Review. No abstract available.

MeSH terms