Correction of photon attenuation and collimator response for a body-contouring SPECT/CT imaging system
- PMID: 15872362
Correction of photon attenuation and collimator response for a body-contouring SPECT/CT imaging system
Abstract
(111)In-Capromab pendetide imaging is indicated for postprostatectomy patients at risk for residual or recurrent disease. However, this study is complicated by relatively long times for tumor uptake and background washout that require imaging to be performed several days after radiopharmaceutical administration. In addition, (111)In-capromab pendetide demonstrates uptake in normal structures that produce images that are interpreted best using correlation with anatomic imaging. Finally, the visual quality of radionuclide imaging can be improved with corrections for photon attenuation and for the geometric response of the radionuclide collimator. Therefore, we have evaluated the advantages of using a commercially available dual-modality SPECT/CT system. In this article, we evaluate a novel iterative reconstruction algorithm using the SPECT/CT data obtained from phantoms and (111)In-capromab pendetide patient studies.
Methods: Phantom data acquired with the dual-head SPECT camera were reconstructed using both filtered backprojection (FBP) and an iterative maximum-likelihood expectation maximization (MLEM) algorithm incorporating corrections for (a) attenuation coefficient at the effective energy of the radionuclide (either (99m)Tc or (111)In) and (b) collimator response based on experimentally measured depth-dependent spatial resolution of the camera. The collimator response model used the coregistered CT image to estimate the source-target distances produced by the patient-contouring logic of the SPECT camera. Spatial resolution was measured using SPECT images of 2 line sources and uniformity from a uniform cylindric tank. Clinical (111)In-capromab pendetide SPECT/CT data were acquired according to the radiopharmaceutical manufacturer's protocol. Region-of-interest (ROI) analysis of a transverse slice at the level of the sacral base produced mean, median, maximum, and minimum counts per pixel for bone marrow and surrounding soft-tissue ROIs. Ratios of the mean capromab pendetide uptake within marrow to uptake within soft tissue were compared for images reconstructed with FBP versus that obtained from the MLEM method with photon attenuation and collimator response corrections.
Results: The source-target distances reconstructed from the patient-specific CT image agreed well with the corresponding values recorded manually from the camera display unit. This information was incorporated into the iterative reconstruction algorithms and improved the quality of SPECT images from phantoms and patients versus SPECT images reconstructed without the depth-dependent collimator response model. Qualitatively, SPECT images reconstructed with corrections for photon attenuation and collimator response showed less background activity and improved target contrast compared with those images reconstructed with FBP. The target-to-background ratio (marrow uptake-to-soft-tissue uptake) was significantly better using MLEM reconstruction than with FBP when mean uptake values were measured.
Conclusion: A priori anatomic data can be used to enhance the quality of the SPECT image when reconstructed using iterative techniques (e.g., MLEM) that use the CT data to produce a patient-specific attenuation map and a collimator response model based on the body contour produced during the SPECT acquisition.
Similar articles
-
A dual modality approach to quantitative quality control in emission tomography.Phys Med Biol. 2005 Aug 7;50(15):N187-94. doi: 10.1088/0031-9155/50/15/N03. Epub 2005 Jul 19. Phys Med Biol. 2005. PMID: 16030376
-
Fast 3D iterative image reconstruction for SPECT with rotating slat collimators.Phys Med Biol. 2009 Feb 7;54(3):715-29. doi: 10.1088/0031-9155/54/3/016. Epub 2009 Jan 9. Phys Med Biol. 2009. PMID: 19131666
-
Progress in SPECT/CT imaging of prostate cancer.Technol Cancer Res Treat. 2006 Aug;5(4):329-36. doi: 10.1177/153303460600500404. Technol Cancer Res Treat. 2006. PMID: 16866563
-
X-ray-based attenuation correction for positron emission tomography/computed tomography scanners.Semin Nucl Med. 2003 Jul;33(3):166-79. doi: 10.1053/snuc.2003.127307. Semin Nucl Med. 2003. PMID: 12931319 Review.
-
Determination of the attenuation map in emission tomography.J Nucl Med. 2003 Feb;44(2):291-315. J Nucl Med. 2003. PMID: 12571222 Review.
Cited by
-
Normal Skeletal Standardized Uptake Values Obtained from Quantitative Single-Photon Emission Computed Tomography/Computed Tomography: Time-Dependent Study on Breast Cancer Patients.Indian J Nucl Med. 2021 Oct-Dec;36(4):398-411. doi: 10.4103/ijnm.ijnm_47_21. Epub 2021 Dec 15. Indian J Nucl Med. 2021. PMID: 35125758 Free PMC article.
-
Investigation of Collimator Influential Parameter on SPECT Image Quality: a Monte Carlo Study.J Biomed Phys Eng. 2015 Mar 4;5(1):39-48. eCollection 2015 Mar. J Biomed Phys Eng. 2015. PMID: 25973410 Free PMC article.
-
Monte carlo study of the effect of collimator thickness on T-99m source response in single photon emission computed tomography.World J Nucl Med. 2012 May;11(2):70-4. doi: 10.4103/1450-1147.103419. World J Nucl Med. 2012. PMID: 23372440 Free PMC article.
-
SPECT-CT: applications in musculoskeletal radiology.Br J Radiol. 2013 Nov;86(1031):20120519. doi: 10.1259/bjr.20120519. Epub 2013 Oct 4. Br J Radiol. 2013. PMID: 24096590 Free PMC article. Review.
-
Brain single-photon emission CT physics principles.AJNR Am J Neuroradiol. 2008 Aug;29(7):1247-56. doi: 10.3174/ajnr.A1175. Epub 2008 Jun 26. AJNR Am J Neuroradiol. 2008. PMID: 18583408 Free PMC article. Review.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous