Comparison of electrically evoked cortical potential thresholds generated with subretinal or suprachoroidal placement of a microelectrode array in the rabbit
- PMID: 15876654
- DOI: 10.1088/1741-2560/2/1/007
Comparison of electrically evoked cortical potential thresholds generated with subretinal or suprachoroidal placement of a microelectrode array in the rabbit
Abstract
The aim of the study was to directly compare the threshold electrical charge density of the retina (retinal threshold) in rabbits for the generation of electrical evoked potentials (EEP) by delivering electrical stimulation with a custom-made microelectrode array (MEA) implanted into either the subretinal or suprachoroidal space. Nine eyes of seven Dutch-belted rabbits were studied. The electroretinogram (ERG), visual evoked potentials (VEP) and EEP were recorded. Electrodes for the VEP and EEP were placed on the dura mater overlying the visual cortex. The EEP was recorded following electrical stimulation of the MEA placed either subretinally beneath the visual streak of the retina or in the suprachoroidal space in the rabbit eye. An ab externo approach was used for placement of the MEA. Liquid perfluorodecaline (PFCL; 0.4 ml) was placed within the vitreous cavity to flatten the neurosensory retina on the MEA after subretinal implantation. The retinal threshold for generation of an EEP was determined for each MEA placement by three consecutive measurements consisting of 100 computer-averaged recordings. Animals were sacrificed at the conclusion of the experiment and the eyes were enucleated for histological examination. The retinal threshold to generate an EEP was 9 +/- 7 nC (0.023 +/- 0.016 mC cm(-2)) within the subretinal space and 150 +/- 122 nC (0.375 +/- 0.306 mC cm(-2)) within the suprachoroidal space. Histology showed disruption of the outer retina with subretinal but not suprachoroidal placement. The retinal threshold to elicit an EEP is significantly lower with subretinal placement of the MEA compared to suprachoroidal placement (P < 0.05). The retinal threshold charge density with a subretinal MEA is well below the published charge limit of 1 mC cm(-2), which is the level below which chronic stimulation of the retina is considered necessary to avoid tissue damage (Shannon 1992 IEEE Trans. Biomed. Eng. 39 424-6).
Similar articles
-
Transretinal electrical stimulation with a suprachoroidal multichannel electrode in rabbit eyes.Jpn J Ophthalmol. 2004 May-Jun;48(3):256-61. doi: 10.1007/s10384-004-0055-1. Jpn J Ophthalmol. 2004. PMID: 15175918
-
Implantation and testing of subretinal film electrodes in domestic pigs.Exp Eye Res. 2006 Feb;82(2):332-40. doi: 10.1016/j.exer.2005.07.007. Epub 2005 Aug 25. Exp Eye Res. 2006. PMID: 16125172
-
Transscleral implantation and neurophysiological testing of subretinal polyimide film electrodes in the domestic pig in visual prosthesis development.J Neural Eng. 2005 Mar;2(1):S57-64. doi: 10.1088/1741-2560/2/1/008. Epub 2005 Feb 22. J Neural Eng. 2005. PMID: 15876656
-
[Research progress of subretinal implant based on electronic stimulation].Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2008 Feb;25(1):212-4, 219. Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2008. PMID: 18435293 Review. Chinese.
-
Electrical Stimulation of the Retina to Produce Artificial Vision.Annu Rev Vis Sci. 2016 Oct 14;2:273-294. doi: 10.1146/annurev-vision-111815-114425. Annu Rev Vis Sci. 2016. PMID: 28532361 Review.
Cited by
-
Acute Rabbit Eye Model for Testing Subretinal Prostheses.Transl Vis Sci Technol. 2019 Oct 2;8(5):20. doi: 10.1167/tvst.8.5.20. eCollection 2019 Sep. Transl Vis Sci Technol. 2019. PMID: 31602345 Free PMC article.
-
Repeated transchoroidal implantation and explantation of compound subretinal prostheses: an exploratory study in rabbits.Jpn J Ophthalmol. 2010 Sep;54(5):467-75. doi: 10.1007/s10384-010-0851-8. Epub 2010 Nov 5. Jpn J Ophthalmol. 2010. PMID: 21052911
-
Inner retinal preservation in rat models of retinal degeneration implanted with subretinal photovoltaic arrays.Exp Eye Res. 2014 Nov;128:34-42. doi: 10.1016/j.exer.2014.09.004. Epub 2014 Sep 16. Exp Eye Res. 2014. PMID: 25224340 Free PMC article.
-
Development of a surgical approach for a wide-view suprachoroidal retinal prosthesis: evaluation of implantation trauma.Graefes Arch Clin Exp Ophthalmol. 2012 Mar;250(3):399-407. doi: 10.1007/s00417-011-1794-6. Epub 2011 Aug 27. Graefes Arch Clin Exp Ophthalmol. 2012. PMID: 21874343
-
Micro/Nano Technologies for High-Density Retinal Implant.Micromachines (Basel). 2019 Jun 22;10(6):419. doi: 10.3390/mi10060419. Micromachines (Basel). 2019. PMID: 31234507 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources