Intensity modulated radiation therapy with multileaf collimators of different leaf widths: a comparison of achievable dose distributions
- PMID: 15878108
- DOI: 10.1016/j.radonc.2005.02.007
Intensity modulated radiation therapy with multileaf collimators of different leaf widths: a comparison of achievable dose distributions
Abstract
Purpose: A planning study to analyze the impact of different leaf widths on the achievable dose distributions with intensity modulated radiation therapy (IMRT).
Methods: Five patients (3 intra- and 2 extra-cranial) with projected planning target volume (PTV) sizes smaller than 10 cm by 10 cm were re-planned with four different multileaf collimators (MLC). Two internal collimators with an isocentric leaf width of 4 and 10 mm and two add-on collimators with an isocentric leaf width of 2.75 and were evaluated. The inverse treatment planning system KonRad (Siemens Medical Solutions) was used to create IMRT 'step & shoot' plans. For each patient the same arrangement of beams and the same parameters for the optimization were used for all MLCs. The beamlet size for all treatment plans was chosen to coincide with the leaf width of the respective MLC. To evaluate the treatment plans 3D dose distributions and dose volume histograms were analyzed. As indicators for the quality of the PTV dose distribution the minimum dose, maximum dose and the standard deviation were used. For the organs at risk (OAR) the equivalent uniform dose (EUD) was calculated. To measure the dose coverage of the PTV the volume (V(90)) that received doses higher than 90% of the prescribed dose was calculated where for the conformity the dose conformity index given by Baltas et al. was determined.
Results: The MLC with the smallest leaf width yields the best mean value of all five patients for the PTV coverage and for the conformity. For the MLCs with the same leaf width, the add-on MLC leads to superior treatment plans than the internal MLC. This is due to the sharper penumbra of the add-on MLC. The number of IMRT field segments to deliver increased by approximately a factor of two if 2. MLC leafs are used instead of the standard 10 mm leafs. In case of the para-spinal patients the EUD value for the spinal cord is only reduced slightly by using MLCs with leaf widths smaller than 5 mm. For the intra-cranial the EUD value for some organs improved with reduced leaf widths while for some organs the 10 mm MLC leafs give comparable values.
Conclusion: As expected the MLC with the smallest leaf width always yields the best PTV coverage. Reducing the leaf width from 4 to 2.75 mm results in a slight enhancement of the PTV coverage. With the selected organ parameters no significant improvement for most OAR was found. The disadvantage of the reduction of the leaf width is the increasing number of segments due to the more complex fluence patterns and therefore an increased delivery time.
Similar articles
-
Impact of IMRT and leaf width on stereotactic body radiotherapy of liver and lung lesions.Int J Radiat Oncol Biol Phys. 2005 Apr 1;61(5):1572-81. doi: 10.1016/j.ijrobp.2004.12.075. Int J Radiat Oncol Biol Phys. 2005. PMID: 15817364
-
Effect of MLC leaf width and PTV margin on the treatment planning of intensity-modulated stereotactic radiosurgery (IMSRS) or radiotherapy (IMSRT).Med Dosim. 2009 Summer;34(2):110-6. doi: 10.1016/j.meddos.2008.06.002. Epub 2008 Jul 1. Med Dosim. 2009. PMID: 19410139
-
Simultaneous beam geometry and intensity map optimization in intensity-modulated radiation therapy.Int J Radiat Oncol Biol Phys. 2006 Jan 1;64(1):301-20. doi: 10.1016/j.ijrobp.2005.08.023. Epub 2005 Nov 14. Int J Radiat Oncol Biol Phys. 2006. PMID: 16289912
-
[Current status and future prospects of 3D treatment planning in radiation therapy, focusing on IMRT].Gan To Kagaku Ryoho. 1999 Aug;26(9):1248-53. Gan To Kagaku Ryoho. 1999. PMID: 10478176 Review. Japanese.
-
Delivery systems of intensity-modulated radiotherapy using conventional multileaf collimators.Med Dosim. 2001 Summer;26(2):169-77. doi: 10.1016/s0958-3947(01)00060-7. Med Dosim. 2001. PMID: 11444519 Review.
Cited by
-
A planning comparison of dynamic IMRT for different collimator leaf thicknesses with helical tomotherapy and RapidArc for prostate and head and neck tumors.Strahlenther Onkol. 2010 Sep;186(9):502-10. doi: 10.1007/s00066-010-2124-3. Epub 2010 Aug 30. Strahlenther Onkol. 2010. PMID: 20803184
-
Optimization of Clarkson's Method for Calculating Absorbed Dose under Compensator Filters used in Intensity-modulated Radiation Therapy.J Biomed Phys Eng. 2020 Oct 1;10(5):575-582. doi: 10.31661/jbpe.v0i0.858. eCollection 2020 Oct. J Biomed Phys Eng. 2020. PMID: 33134216 Free PMC article.
-
Implications of a high-definition multileaf collimator (HD-MLC) on treatment planning techniques for stereotactic body radiation therapy (SBRT): a planning study.Radiat Oncol. 2009 Jul 10;4:22. doi: 10.1186/1748-717X-4-22. Radiat Oncol. 2009. PMID: 19591687 Free PMC article.
-
Dosimetric effect of multileaf collimator leaf width in intensity-modulated radiotherapy delivery techniques for small- and large-volume targets.J Med Phys. 2011 Apr;36(2):72-7. doi: 10.4103/0971-6203.79690. J Med Phys. 2011. PMID: 21731222 Free PMC article.
-
Optimization of Minimum Segment Width Parameter in the Intensity-Modulated Radiotherapy Plan for Esophageal Cancer.Int J Gen Med. 2021 Dec 16;14:9913-9921. doi: 10.2147/IJGM.S336269. eCollection 2021. Int J Gen Med. 2021. PMID: 34938110 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials