Wet radio-frequency ablation using multiple electrodes: comparative study of bipolar versus monopolar modes in the bovine liver

Affiliations

Wet radio-frequency ablation using multiple electrodes: comparative study of bipolar versus monopolar modes in the bovine liver

Jeong Min Lee et al. Eur J Radiol. 2005 Jun.

Abstract

Objective: To determine whether hypertonic saline (HS)-mediated bipolar radio-frequency (rf) application as advantages over monopolar simultaneous and alternating rf applications for creating larger areas of coagulation necrosis.

Materials and methods: A total of 60 rf ablations using double perfused-cooled electrodes and a 200 W generator (CC-3 model, Radionics) were performed in three different modes in explanted bovine livers: simultaneous monopolar mode (groups A and A'); alternating monopolar mode (groups B and B'); or bipolar mode (groups C and C'). Electrodes were placed at inter-electrode distances of 3 and 5 cm, and HS (6% NaCl solution) was instilled into tissue at a rate of 1 mL/min through the electrodes. rf was applied for 10 (3 cm distance) or 15 min (5 cm distance). During rf application, we measured the tissue temperature at the mid-point between the two electrodes. Dimensions of the thermal ablation zones, and temperatures were compared between the 3 groups using analysis of variance or the Kruskal-Wallis test. To compare configurations of the ablation zones in each group, the ratio of longitudinal diameter (Dl) to vertical diameter (D(v)) was calculated.

Results: With a 3-cm inter-electrode spacing, the D(v) between the electrodes of ablated lesions was 2.4 +/- 1.2 cm in group A, 4.5 +/- 1.0 cm in group B, and 6.1 +/- 0.9 cm in group C (P < 0.05), and at a 5-cm spacing, groups B' and C' produced a single ablation area, but group A' produced two separated ablation spheres: the D(v)s were 1.4 +/- 0.2 in group A, 2.9 +/- 1.0 mm in group B, and 6.6 +/- 0.4 cm in group C (P < 0.05). For both 5- and 3-cm spacings, the temperatures at the mid-point were higher in bipolar mode than in either monopolar simultaneous or alternating modes. The ratios of Dl/D(v) of groups A, B, and C were 2.5 +/- 0.2, 1.4 +/- 0.1, and 1.1 +/- 0.1, respectively, and the corresponding figures of groups A', B' and C' were 4.5 +/- 0.2, 2.7 +/- 0.1, and 1.1 +/- 0.1, respectively (P < 0.05).

Conclusion: HS-enhanced bipolar rf ablation creates larger, more regular coagulation necrosis than either monopolar simultaneous or alternating rf ablation.