The first use of live continuous glucose monitoring in patients on extracorporeal life support
Abstract
Background: Evidence suggests that glycemic control provides clinical benefit to critically ill patients. The Extracorporeal Glucose Monitoring System (EGMS), Medtronic Minimed, Northridge, CA) has been developed to measure real-time, continuous blood glucose concentrations in patients on extracorporeal bypass. This pilot study reports the first in vitro and in vivo evaluations of EGMS in an extracorporeal circuit.
Methods: In an in vitro study, three EGMS sensors were inserted in a neonatal extracorporeal circuit. Circuit blood glucose levels were altered by saline dilution and dextrose infusion. EGMS sensors were then inserted into the venous return limb of the extracorporeal circuit in a cohort of six critically ill infants on extracorporeal life support (ECLS). Continuous glucose measurements were compared with laboratory and bedside glucose values at predefined intervals. Linear regression analyses and the Clarke error grid were constructed to analyze device accuracy.
Results: All three in vitro EGMS sensors recorded real-time data without interruption for 48 h. EGMS glucose measurements closely correlated with reference levels (R (2) = 0.93). EGMS glucose values demonstrated an approximate 7-10 min lag while glucose concentrations were rapidly changing. Eight EGMS devices were inserted into six neonates on ECLS on day of life 6 +/- 3. EGMS correlated well with laboratory glucose (R2 = 0.61) and bedside glucose during a hyperinsulinemic euglycemic clamp (R2 = 0.78). On the Clarke error grid, 98% of readings were within zones A and B using laboratory glucose as reference, and 100% were within zones A and B using bedside glucose measurements. Blood glucose range during the in vitro study was 19-295 mg/dL and during the in vivo study was 80-257 mg/dL.
Conclusions: This pilot study suggests that EGMS is a reliable tool for measuring continuous blood glucose in critically ill patients connected to an extracorporeal circuit, although important limitations exist. Potential applications of this technology include intensive glucose monitoring in patients on ECLS, cardiopulmonary bypass, and renal replacement therapy.
Similar articles
-
Use of a continuous glucose sensor in an extracorporeal life support circuit.J Diabetes Sci Technol. 2011 Jan 1;5(1):93-8. doi: 10.1177/193229681100500113. J Diabetes Sci Technol. 2011. PMID: 21303630 Free PMC article.
-
Experience with the continuous glucose monitoring system in a medical intensive care unit.Diabetes Technol Ther. 2004 Jun;6(3):339-47. doi: 10.1089/152091504774198034. Diabetes Technol Ther. 2004. PMID: 15198837
-
Evaluating the accuracy, reliability, and clinical applicability of continuous glucose monitoring (CGM): Is CGM ready for real time?Diabetes Technol Ther. 2009 Jan;11(1):11-8. doi: 10.1089/dia.2008.0041. Diabetes Technol Ther. 2009. PMID: 19132850
-
The public health impact of the MiniMed Continuous Glucose Monitoring System (CGMS)-an assessment of the literature.Diabetes Technol Ther. 2004 Aug;6(4):518-22. doi: 10.1089/1520915041705956. Diabetes Technol Ther. 2004. PMID: 15321009 Review.
-
Continuous glucose monitoring with glucose sensors: calibration and assessment criteria.Diabetes Technol Ther. 2003;5(4):572-86. doi: 10.1089/152091503322250596. Diabetes Technol Ther. 2003. PMID: 14511412 Review.
Cited by
-
Continuous glucose monitors prove highly accurate in critically ill children.Crit Care. 2010;14(5):R176. doi: 10.1186/cc9280. Epub 2010 Oct 6. Crit Care. 2010. PMID: 20925914 Free PMC article.
-
Use of a continuous glucose sensor in an extracorporeal life support circuit.J Diabetes Sci Technol. 2011 Jan 1;5(1):93-8. doi: 10.1177/193229681100500113. J Diabetes Sci Technol. 2011. PMID: 21303630 Free PMC article.
-
Recent progress in mechanical artificial pancreas.J Artif Organs. 2009;12(3):141-9. doi: 10.1007/s10047-009-0463-6. Epub 2009 Sep 19. J Artif Organs. 2009. PMID: 19894087 Review.
-
Hypoglycemia in critically ill children.J Diabetes Sci Technol. 2012 Jan 1;6(1):48-57. doi: 10.1177/193229681200600107. J Diabetes Sci Technol. 2012. PMID: 22401322 Free PMC article.
-
Blood-induced interference of glucose sensor function in vitro: implications for in vivo sensor function.J Diabetes Sci Technol. 2007 Nov;1(6):842-9. doi: 10.1177/193229680700100607. J Diabetes Sci Technol. 2007. PMID: 19885155 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical