Response to nocturnal alarms using a real-time glucose sensor
- PMID: 15929675
- PMCID: PMC1482828
- DOI: 10.1089/dia.2005.7.440
Response to nocturnal alarms using a real-time glucose sensor
Abstract
Background: The objective of this study was to determine how subjects responded to alarms for hypo- and hyperglycemia while they were sleeping.
Research design and methods: Twenty subjects with type 1 diabetes (4-17 years old) were admitted to a clinical research center for approximately 24 h. Each subject wore two GlucoWatch G2 Biographers (GW2B) (Cygnus, Inc., Redwood City, CA) and was videotaped using an infrared camera from 9 p.m. to 7 a.m. The videotapes were reviewed to determine if the GW2B alarms were audible on the tape and to document the subject's response to the alarms. Because many alarms can occur surrounding a change in blood glucose, GW2B alarm "events" are defined as a one or more alarms separated from previous alarms by more than 30 min.
Results: Downloaded data from the biographers identified 240 individual alarms, 75% of which occurred while the subject was sleeping. Of the 240 alarms 68% were audible on the videotape. Subjects awoke to 29% of individual alarms and to 66% of alarm events. Subjects 4-6 years old responded to 17% of alarms, 7-11 year olds responded to 20% of alarms, adolescents responded to 53% of alarms, and parents responded to 37% of alarms. Subjects awoke to 40% of the first alarm during the night, but to only 28% of subsequent alarms. There were 11 events when the glucose was confirmed to be < or = 70 mg/dL, and in each case the subject was awoken. Fifty-five percent of alarm events occurred when there was no hypo- or hyperglycemia confirmed by a reference glucose value.
Conclusions: Subjects awoke to 29% of individual alarms and to 66% of alarm events. Subjects awoke during all alarm events when hypoglycemia was confirmed, but there was a high incidence of false alarms.
Figures

Similar articles
-
GlucoWatch G2 Biographer alarm reliability during hypoglycemia in children.Diabetes Technol Ther. 2004 Oct;6(5):559-66. doi: 10.1089/dia.2004.6.559. Diabetes Technol Ther. 2004. PMID: 15628809 Free PMC article.
-
Long-Term Home Study on Nocturnal Hypoglycemic Alarms Using a New Fully Implantable Continuous Glucose Monitoring System in Type 1 Diabetes.Diabetes Technol Ther. 2015 Nov;17(11):780-6. doi: 10.1089/dia.2014.0375. Epub 2015 Jul 15. Diabetes Technol Ther. 2015. PMID: 26177299
-
Use of the Cygnus GlucoWatch biographer at a diabetes camp.Pediatrics. 2004 Jan;113(1 Pt 1):108-11. doi: 10.1542/peds.113.1.108. Pediatrics. 2004. PMID: 14702457 Clinical Trial.
-
Continuous glucose monitoring: physiological and technological challenges.Curr Diabetes Rev. 2008 Aug;4(3):175-80. doi: 10.2174/157339908785294488. Curr Diabetes Rev. 2008. PMID: 18690898 Review.
-
Reduction of number and duration of hypoglycemic events by glucose prediction methods: a proof-of-concept in silico study.Diabetes Technol Ther. 2013 Jan;15(1):66-77. doi: 10.1089/dia.2012.0208. Diabetes Technol Ther. 2013. PMID: 23297671 Review.
Cited by
-
Hypoglycaemia in type 1 diabetes: technological treatments, their limitations and the place of psychology.Diabetologia. 2018 Apr;61(4):761-769. doi: 10.1007/s00125-018-4566-6. Epub 2018 Feb 8. Diabetologia. 2018. PMID: 29423581 Free PMC article. Review.
-
Real-time continuous glucose monitoring significantly reduces severe hypoglycemia in hypoglycemia-unaware patients with type 1 diabetes.Diabetes Care. 2013 Dec;36(12):4160-2. doi: 10.2337/dc13-0939. Epub 2013 Oct 8. Diabetes Care. 2013. PMID: 24103902 Free PMC article.
-
"Turn it off!": diabetes device alarm fatigue considerations for the present and the future.J Diabetes Sci Technol. 2013 May 1;7(3):789-94. doi: 10.1177/193229681300700324. J Diabetes Sci Technol. 2013. PMID: 23759412 Free PMC article.
-
Predictive Low-Glucose Insulin Suspension Reduces Duration of Nocturnal Hypoglycemia in Children Without Increasing Ketosis.Diabetes Care. 2015 Jul;38(7):1197-204. doi: 10.2337/dc14-3053. Epub 2015 Jun 6. Diabetes Care. 2015. PMID: 26049549 Free PMC article. Clinical Trial.
-
Increasing the low-glucose alarm of a continuous glucose monitoring system prevents exercise-induced hypoglycemia without triggering any false alarms.Diabetes Care. 2011 Jun;34(6):e109. doi: 10.2337/dc10-2243. Diabetes Care. 2011. PMID: 21617102 Free PMC article. Clinical Trial. No abstract available.
References
-
- Marrero DG, Guare JC, Vandagriff JL, Fineberg NS. Fear of hypoglycemia in the parents of children and adolescents with diabetes: maladaptive or healthy response? Diabetes Educ. 1997;23(3):281–6. - PubMed
-
- Davis EA, Keating B, Byrne GC, Russell M, Jones TW. Hypoglycemia: incidence and clinical predictors in a large population-based sample of children and adolescents with IDDM. Diabetes Care. 1997;20(1):22–5. - PubMed
-
- Rovet JF. The effect of hypoglycemic seizures on cognitive function in children with diabetes: a 7-year prospective study. J. Pediatr. 1999;134:503–506. - PubMed
-
- Sovik O, Thordarson H. Dead-in-bed syndrome in young diabetic patients. Diabetes Care. 1999;22 (Suppl 2):B40–2. - PubMed
-
- Potts RO, Tamada JA, Tierney MJ. Glucose monitoring by reverse iontophoresis. Diabetes Metab Res Rev. 2002;18 (Suppl 1):S49–53. - PubMed
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Medical