Home blood glucose prediction: validation, safety, and efficacy testing in clinical diabetes
- PMID: 15929680
- DOI: 10.1089/dia.2005.7.487
Home blood glucose prediction: validation, safety, and efficacy testing in clinical diabetes
Abstract
Background: Patients with diabetes do daily self-monitoring of blood glucose (SMBG). For such patients, we devised an engine that predicts not only the expected blood glucose level at the next meal but also the pending risks of hypoglycemia. The purpose of this study was to validate the predictions and provide evidence of the safety and efficacy of using predicted data in dosing decision support for routine patient care.
Research design and methods: The prediction engine is a computer program that accesses a central database into which daily records of self-monitored blood glucose data are captured by direct access either across the WWW or by an interactive voice response service on-line 24/7. Validation was done by comparison of predicted values to the subsequently observed data using a Clarke Error Grid. Safety focused on body weight and the frequency of hypoglycemia. Efficacy was judged according to glycated hemoglobin and daily insulin dosages. The experimental design contrasted patients in the tight control (TC) group who had been recently converted to intensified (basal-bolus) therapy with patients in the poor control (PC) group on conventional therapy and who were referred to begin intensified therapy. Both groups accessed the remote database to report their daily SMBG. Predicted glucose values were used in dosing decision support for the PC but not the TC group.
Results: Over the 6-month study period a total of 30,129 blood glucose levels were reported by the 54 study patients, and some 24,953 blood glucose predictions were made. Of these, 83% were in the clinically acceptable zones of the Clarke Error Grid. When these data were used for dosing decision support in the PC group, glycated hemoglobin levels fell significantly from 9.7 +/- 1.7% to 7.9 +/- 1.2%, and hypoglycemia dropped fourfold. Total daily insulin doses declined 22%, while body weight remained constant. In the parallel TC group (n = 24), glycated hemoglobin also fell, but only slightly from 7.6 +/- 0.9% to 7.2 +/- 1.1%, while daily insulin doses, rates of hypoglycemia and body weight remained constant.
Conclusions: A novel engine is capable of generating meaningful predictions of blood glucose levels. Use of these validated predictions in decision support for managing medication doses proved safe and efficacious.
Similar articles
-
Home blood glucose prediction: clinical feasibility and validation in islet cell transplantation candidates.Diabetologia. 2005 Jul;48(7):1273-9. doi: 10.1007/s00125-005-1805-4. Epub 2005 Jun 3. Diabetologia. 2005. PMID: 15933858
-
A graphical user interface for diabetes management that integrates glucose prediction and decision support.Diabetes Technol Ther. 2005 Apr;7(2):264-73. doi: 10.1089/dia.2005.7.264. Diabetes Technol Ther. 2005. PMID: 15857228
-
Information technology and home glucose clamping.Diabetes Technol Ther. 2001 Fall;3(3):377-86. doi: 10.1089/15209150152607150. Diabetes Technol Ther. 2001. PMID: 11762516
-
Diabetes technology and treatments in the paediatric age group.Int J Clin Pract Suppl. 2011 Feb;(170):76-82. doi: 10.1111/j.1742-1241.2010.02582.x. Int J Clin Pract Suppl. 2011. PMID: 21323816 Review.
-
Understanding error grid analysis.Diabetes Care. 1997 Jun;20(6):911-2. doi: 10.2337/diacare.20.6.911. Diabetes Care. 1997. PMID: 9167098 Review. No abstract available.
Cited by
-
Home blood glucose prediction: clinical feasibility and validation in islet cell transplantation candidates.Diabetologia. 2005 Jul;48(7):1273-9. doi: 10.1007/s00125-005-1805-4. Epub 2005 Jun 3. Diabetologia. 2005. PMID: 15933858
-
Predicting postprandial glucose excursions using gaussian process regression.J Diabetes Sci Technol. 2009 Mar 1;3(2):405-7. doi: 10.1177/193229680900300226. J Diabetes Sci Technol. 2009. PMID: 20144374 Free PMC article. No abstract available.
-
A novel method for measuring patients' adherence to insulin dosing guidelines: introducing indicators of adherence.BMC Med Inform Decis Mak. 2008 Dec 5;8:55. doi: 10.1186/1472-6947-8-55. BMC Med Inform Decis Mak. 2008. PMID: 19061492 Free PMC article.
-
Prescription checking device promises to resolve intractable hypoglycemia.J Diabetes Sci Technol. 2009 May 1;3(3):524-32. doi: 10.1177/193229680900300317. J Diabetes Sci Technol. 2009. PMID: 20144291 Free PMC article.
-
Recommending blood glucose monitors, a pharmacy perspective.J Diabetes Sci Technol. 2007 Mar;1(2):228-30. doi: 10.1177/193229680700100213. J Diabetes Sci Technol. 2007. PMID: 19888411 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical