Statistical error in isothermal titration calorimetry: variance function estimation from generalized least squares
- PMID: 15936713
- DOI: 10.1016/j.ab.2005.04.026
Statistical error in isothermal titration calorimetry: variance function estimation from generalized least squares
Abstract
The method of generalized least squares (GLS) is used to assess the variance function for isothermal titration calorimetry (ITC) data collected for the 1:1 complexation of Ba(2+) with 18-crown-6 ether. In the GLS method, the least squares (LS) residuals from the data fit are themselves fitted to a variance function, with iterative adjustment of the weighting function in the data analysis to produce consistency. The data are treated in a pooled fashion, providing 321 fitted residuals from 35 data sets in the final analysis. Heteroscedasticity (nonconstant variance) is clearly indicated. Data error terms proportional to q(i) and q(i)/v are well defined statistically, where q(i) is the heat from the ith injection of titrant and v is the injected volume. The statistical significance of the variance function parameters is confirmed through Monte Carlo calculations that mimic the actual data set. For the data in question, which fall mostly in the range of q(i)=100-2000 microcal, the contributions to the data variance from the terms in q(i)(2) typically exceed the background constant term for q(i)>300 microcal and v<10 microl. Conversely, this means that in reactions with q(i) much less than this, heteroscedasticity is not a significant problem. Accordingly, in such cases the standard unweighted fitting procedures provide reliable results for the key parameters, K and DeltaH(degrees) and their statistical errors. These results also support an important earlier finding: in most ITC work on 1:1 binding processes, the optimal number of injections is 7-10, which is a factor of 3 smaller than the current norm. For high-q reactions, where weighting is needed for optimal LS analysis, tips are given for using the weighting option in the commercial software commonly employed to process ITC data.
Similar articles
-
Calorimetric vs. van't Hoff binding enthalpies from isothermal titration calorimetry: Ba2+-crown ether complexation.Biophys Chem. 2004 Jul 1;110(1-2):15-24. doi: 10.1016/j.bpc.2003.12.011. Biophys Chem. 2004. PMID: 15223140
-
Weighted least squares in calibration: the problem with using "quality coefficients" to select weighting formulas.J Chromatogr B Analyt Technol Biomed Life Sci. 2008 Sep 1;872(1-2):162-6. doi: 10.1016/j.jchromb.2008.07.043. Epub 2008 Aug 6. J Chromatogr B Analyt Technol Biomed Life Sci. 2008. PMID: 18706869
-
Van't Hoff analysis of K degrees (T): how good...or bad?Biophys Chem. 2006 Mar 20;120(2):114-20. doi: 10.1016/j.bpc.2005.10.012. Epub 2005 Nov 21. Biophys Chem. 2006. PMID: 16303233
-
Advances in the analysis of isothermal titration calorimetry data for ligand-DNA interactions.Methods. 2007 Jun;42(2):162-72. doi: 10.1016/j.ymeth.2007.01.010. Methods. 2007. PMID: 17472898 Review.
-
Isothermal titration calorimetry: experimental design, data analysis, and probing macromolecule/ligand binding and kinetic interactions.Methods Cell Biol. 2008;84:79-113. doi: 10.1016/S0091-679X(07)84004-0. Methods Cell Biol. 2008. PMID: 17964929 Review.
Cited by
-
Preferential binding of K+ ions in the selectivity filter at equilibrium explains high selectivity of K+ channels.J Gen Physiol. 2012 Dec;140(6):671-9. doi: 10.1085/jgp.201210855. Epub 2012 Nov 12. J Gen Physiol. 2012. PMID: 23148260 Free PMC article.
-
Entropy-enthalpy compensation: role and ramifications in biomolecular ligand recognition and design.Annu Rev Biophys. 2013;42:121-42. doi: 10.1146/annurev-biophys-083012-130318. Annu Rev Biophys. 2013. PMID: 23654303 Free PMC article. Review.
-
Isothermal titration calorimetry of RNA.Methods. 2009 Mar;47(3):198-205. doi: 10.1016/j.ymeth.2008.09.003. Epub 2008 Oct 7. Methods. 2009. PMID: 18835447 Free PMC article. Review.
-
Fitting two- and three-site binding models to isothermal titration calorimetric data.Methods. 2015 Apr;76:124-136. doi: 10.1016/j.ymeth.2014.11.018. Epub 2014 Dec 5. Methods. 2015. PMID: 25484338 Free PMC article.
-
Uncertainty in protein-ligand binding constants: asymmetric confidence intervals versus standard errors.Eur Biophys J. 2021 May;50(3-4):661-670. doi: 10.1007/s00249-021-01518-4. Epub 2021 Apr 10. Eur Biophys J. 2021. PMID: 33837826
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous