Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Clinical Trial
. 2005 Jun 22;102(1):95-102.
doi: 10.1016/j.ijcard.2004.04.001.

The SV stent study: a prospective, multicentre, angiographic evaluation of the BiodivYsio phosphorylcholine coated small vessel stent in small coronary vessels

Affiliations
Clinical Trial

The SV stent study: a prospective, multicentre, angiographic evaluation of the BiodivYsio phosphorylcholine coated small vessel stent in small coronary vessels

A Bakhai et al. Int J Cardiol. .

Abstract

Objective: To evaluate the use of the phosphorylcholine (PC) coated BiodivYsio small vessel (SV) stent in native coronary vessels of small calibre.

Design and setting: Prospective, multi-centre, multi-national registry with 6-month clinical and core-lab angiographic follow-up. Adverse events were adjudicated by a Clinical Events Committee (CEC) and included peri-procedural analysis of cardiac enzymes.

Patients: Patients with signs or symptoms of ischaemia with an identified target lesion in an epicardial vessel with reference diameter 2.0-2.75 mm were enrolled. Intervention in other epicardial territories in the same patient was permitted.

Results: Recruitment of 150 consecutive lesions (in 143 patients) was completed in 19 centres in Europe and Israel. The stent was deployed successfully in all but one lesion. At 6 months, 1 patient (1%) had experienced sudden cardiac death, 4 further patients (3%) had a non-Q wave MI, and a further 24 patients (17%) had repeat revascularisation of a study target vessel. The mean reference vessel diameter prior to stenting was 2.2 mm (S.D. 0.4). Mean minimal luminal diameters at pre-procedure, post procedure and follow-up were 0.6 mm (S.D. 0.3), 2.0 mm (S.D. 0.4) and 1.2 mm (S.D. 0.6), respectively. The late lumen loss index was 0.55 (S.D. 0.53) with a binary restenosis rate of 32%.

Conclusions: In stenting of selected lesions in small vessels, the BiodivYsio SV stent demonstrated high rates of implant success. The rates of major adverse cardiac events (MACE), angiographic restenosis and repeat revascularisation are similar to those reported in other small vessel bare metal stent studies.