Biocompatibility screening in cardiovascular implants
- PMID: 15940438
- DOI: 10.1007/s00392-005-0231-4
Biocompatibility screening in cardiovascular implants
Abstract
Background: Interest in information on biocompatibility of implants is increasing. The purpose of this paper is to discuss methods and results of pathological biocompatibility screening of explanted cardiovascular implants.
Methods: Use of standard histology after embedding in paraffin is limited since metallic implants have to be removed during workup with disruption of the specimen. Alternatively, tissue blocks containing an implant can be embedded in methylmethacrylate or hydroxyethylmethacrylate and processed by sectioning with a diamond cutter and grinding, thus leaving the implant in situ and saving the tissue/implant interface for detection of local inflammatory reactions. Another important aspect of evaluation is the progress of thrombus organisation after initial fibrin clotting on the metal surface or in the inner part of occlusion devices. New methacrylate resins and embedding techniques allow for specific immunohistochemical staining of the specimen thus enabling characterisation of tissues surrounding the implant. Information on endothelialisation of the vascular surface of the implant can be obtained by means of immunohistochemistry or by scanning electron microscopy.
Results: Illustrating the use of these technologies, we demonstrate findings in tissue specimens from animal studies with different types of devices (i.e. stents, occlusion devices). We present corresponding findings in human specimens with implants that were removed during corrective surgery for congenital heart defects. Early endothelialisation of the vascular surface was seen after implantation in all types of devices. Cells within occlusion devices could be characterised histologically and immunohistochemically as fibromuscular cells as seen in intimal hyperplasia after stent implantation. Inflammatory implant-host reactions ranged from mild to moderate (medical grade stainless steel, nitinol) to severe (polytetrafluoroethylene [PTFE]).
Conclusions: With an optimal work-up of cardiovascular implants, ingrowth and endothelialisation as well as inflammatory reactions in the surrounding tissue can be assessed. This information allows evaluation of individual tissue reactions to the implant and may serve as valuable basis for optimisation of biocompatibility by implant modification.
Similar articles
-
Immunohistochemical characterization of neotissues and tissue reactions to septal defect-occlusion devices.Circ Cardiovasc Interv. 2009 Apr;2(2):90-6. doi: 10.1161/CIRCINTERVENTIONS.108.810507. Epub 2009 Feb 20. Circ Cardiovasc Interv. 2009. PMID: 20031701
-
Biocompatibility of septal defect closure devices.Heart. 2007 Apr;93(4):444-9. doi: 10.1136/hrt.2006.098103. Epub 2006 Oct 11. Heart. 2007. PMID: 17035510 Free PMC article.
-
Evaluation of in vivo biocompatibility of different devices for interventional closure of the patent ductus arteriosus in an animal model.Heart. 2000 May;83(5):570-3. doi: 10.1136/heart.83.5.570. Heart. 2000. PMID: 10768911 Free PMC article.
-
[Classification of prosthetic loosening and determination of wear particles].Pathologe. 2008 Nov;29 Suppl 2:232-9. doi: 10.1007/s00292-008-1070-7. Pathologe. 2008. PMID: 18820921 Review. German.
-
Non-invasive Characterization of Immune Responses to Biomedical Implants.Ann Biomed Eng. 2016 Mar;44(3):693-704. doi: 10.1007/s10439-015-1470-9. Epub 2015 Oct 5. Ann Biomed Eng. 2016. PMID: 26438448 Review.
Cited by
-
Neurosurgical Approaches to Brain Tissue Harvesting for the Establishment of Cell Cultures in Neural Experimental Cell Models.Materials (Basel). 2021 Nov 13;14(22):6857. doi: 10.3390/ma14226857. Materials (Basel). 2021. PMID: 34832259 Free PMC article. Review.
-
Secondary residual shunt after atrial septal defect closure with an amplatzer occluder: surgical removal and evaluation of device biocompatibility after 7 years.Pediatr Cardiol. 2010 Oct;31(7):1107-10. doi: 10.1007/s00246-010-9757-y. Epub 2010 Aug 1. Pediatr Cardiol. 2010. PMID: 20680260
-
Role of animal models for percutaneous atrial septal defect closure.J Thorac Dis. 2018 Sep;10(Suppl 24):S2966-S2974. doi: 10.21037/jtd.2018.07.119. J Thorac Dis. 2018. PMID: 30305957 Free PMC article. Review.
-
Swelling characteristics and biocompatibility of ionic liquid based hydrogels for biomedical applications.PLoS One. 2020 Apr 20;15(4):e0231421. doi: 10.1371/journal.pone.0231421. eCollection 2020. PLoS One. 2020. PMID: 32310981 Free PMC article.
-
Biomaterials and host versus graft response: a short review.Bosn J Basic Med Sci. 2016 Feb 19;16(2):82-90. doi: 10.17305/bjbms.2016.525. Bosn J Basic Med Sci. 2016. PMID: 26894284 Free PMC article. Review.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources