Abstract
The movements of the upper cervical spine were measured by fluoroscopy in 20 patients during laryngoscopy with the Bonfils intubation fibrescope and the Macintosh laryngoscope. Laryngoscopy with both the Bonfils intubation fibrescope and the Macintosh laryngoscope resulted in significant extension of the cervical spine as compared to the neutral position but this extension was significantly less with the Bonfils intubation fibrescope than with the Macintosh (p = 0.001). However, the atlanto-occipital distance was significantly greater during laryngoscopy with the Bonfils intubation fibrescope (p = 0.002), and the angle between the occiput and C1 differed significantly between the two techniques (p = 0.001). With the Bonfils intubation fibrescope, significantly less extension was also found at the C1/C2 and C3/C4 levels (p = 0.001 and p = 0.049, respectively). There is therefore significantly less movement of the upper cervical spine during laryngoscopy with the Bonfils fibrescope compared with the Macintosh laryngoscope.
Similar articles
-
[Cervical spine movement during orotracheal intubation using the McCoy laryngoscope compared with the Macintosh and the Miller laryngoscopes].Masui. 1997 Jan;46(1):124-7. Masui. 1997. PMID: 9028096 Japanese.
-
Three-dimensional cervical spine movement during intubation using the Macintosh and Bullard laryngoscopes, the bonfils fibrescope and the intubating laryngeal mask airway.Eur J Anaesthesiol. 2004 Nov;21(11):907-13. doi: 10.1017/s0265021504000274. Eur J Anaesthesiol. 2004. PMID: 15717709 Clinical Trial.
-
Cervical spine movement during laryngoscopy using the Airway Scope compared with the Macintosh laryngoscope.Anaesthesia. 2007 Oct;62(10):1050-5. doi: 10.1111/j.1365-2044.2007.05188.x. Anaesthesia. 2007. PMID: 17845658 Clinical Trial.
-
Clinical uses of the Bonfils Retromolar Intubation Fiberscope: a review.Anesth Analg. 2012 Oct;115(4):855-66. doi: 10.1213/ANE.0b013e318265bae2. Epub 2012 Sep 5. Anesth Analg. 2012. PMID: 22956530 Review.
-
Cervical spine movements during laryngoscopy and orotracheal intubation: a systematic review and meta-analysis.Emerg Med J. 2023 Apr;40(4):300-307. doi: 10.1136/emermed-2021-211160. Epub 2022 Oct 31. Emerg Med J. 2023. PMID: 36316103
Cited by
-
A comparative randomized trial of intubation success in difficult intubation cases: the use of a Frova intubation catheter versus a Bonfils intubation fiberoscope.Wideochir Inne Tech Maloinwazyjne. 2019 Dec;14(4):486-494. doi: 10.5114/wiitm.2019.83610. Epub 2019 Mar 14. Wideochir Inne Tech Maloinwazyjne. 2019. PMID: 31908693 Free PMC article.
-
Intubation during spinal motion restriction using the LuboTM cervical collar - a manikin simulation study.Afr J Emerg Med. 2022 Dec;12(4):327-332. doi: 10.1016/j.afjem.2022.06.009. Epub 2022 Jul 26. Afr J Emerg Med. 2022. PMID: 35919101 Free PMC article.
-
Comparison of the Airtraq and Truview laryngoscopes to the Macintosh laryngoscope for use by Advanced Paramedics in easy and simulated difficult intubation in manikins.BMC Emerg Med. 2009 Feb 13;9:2. doi: 10.1186/1471-227X-9-2. BMC Emerg Med. 2009. PMID: 19216776 Free PMC article. Clinical Trial.
-
Bonfils intubation fibrescope: use in simulation-based intubation training for medical students in comparison to MacIntosh laryngoscope.BMC Res Notes. 2016 Feb 27;9:127. doi: 10.1186/s13104-016-1937-2. BMC Res Notes. 2016. PMID: 26920895 Free PMC article.
-
AIRWAY MANAGEMENT WITH RIGID BRONCHOSCOPE IN ADULT PATIENT DURING CERVICAL SPINE SURGERY: A CASE REPORT.Acta Clin Croat. 2023 Apr;62(Suppl1):142-148. doi: 10.20471/acc.2023.62.s1.19. Acta Clin Croat. 2023. PMID: 38746614 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous