. 2005 Mar;88(3):340-4.

Tonometry after laser in situ keratomileusis treatment: a preliminary study in Thai patients

Affiliations
  • PMID: 15962641

Tonometry after laser in situ keratomileusis treatment: a preliminary study in Thai patients

Ngamkae Ruangvaravate et al. J Med Assoc Thai. 2005 Mar.

Abstract

Objective: To evaluate the change in intraocular pressure (IOP) measurement by Goldmann applanation tonometer after Laser in Situ Keratomileusis (LASIK) for myopia and myopic astigmatism, and to assess the correlation between the changes of lOP reading and the reduction of central corneal thickness (CCT) after LASIK in Thai patients.

Study design: Prospective correlational study.

Material and method: LASIK was performed on 65 eyes of 33 patients for correction of myopia and myopic astigmatism. IOP was measured by Goldmann applanation tonometer before and 3 months after LASIK. The correlation between the change in IOP reading and the change in central corneal thickness were evaluated.

Results: IOP reading was significantly reduced by mean of 2.9 +/- 2.5 mmHg (p = 0.0001). The authors used Pearson analysis to study the correlation between the change in IOP and the reduction of CCT. In subgroup analysis the patients were divided by degree of myopia: group 1, myopia less than -3 diopters (D) (n = 14); group 2, myopia -3 to -6 D (n = 31); group 3, myopia greater than -6 D (n = 20)). The result showed more correlation in higher myopia group (Pearson; r = 0.158 in group 3, r = -0.098 in group 2 and r = -0.102 in group 1).

Conclusion: Goldmann applanation tonometry underestimates the IOP in thin cornea. Variability in CCT is a potent confounder of this tonometry technique. Therefore, it has important implications for considering CCT measurement incorporated with Goldmann applanation tonometry for glaucoma diagnosis especially in myopic patients who undergo LASIK surgery.