The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry
- PMID: 15965165
- PMCID: PMC1772720
- DOI: 10.1136/bjo.2004.056622
The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry
Abstract
Aims: To evaluate the influence of central corneal thickness (CCT) on intraocular pressure (IOP) measurements made with the Goldmann applanation tonometer (GAT), Tono-Pen XL, ocular blood flow tonograph (OBF), and Canon TX-10 non-contact tonometer (NCT).
Methods: CCT was recorded for either eye (randomly selected) of each of 105 untreated patients with ocular hypertension and glaucoma attending the glaucoma research unit at Moorfields Eye Hospital. For each of the selected eyes, IOP was measured with the GAT (two observers), Tono-Pen, OBF, and NCT in a randomised order. The relation of measured IOP and of inter-tonometer differences with CCT and subject age was explored by linear regression analysis.
Results: A significant association between measured IOP and CCT was found with each instrument. The change in measured IOP for a 10 mum increase in CCT was 0.28, 0.31, 0.38, and 0.46 for the GAT, Tono-Pen, OBF, and NCT, respectively (all p< or = 0.05). There was a significant association between the NCT/GAT differences and CCT, with a tendency of NCT to overestimate GAT in eyes with thicker corneas. There was a significant association between GAT/Tono-Pen and OBF/Tono-Pen differences and age, with a tendency of GAT and OBF to overestimate the Tono-Pen in eyes of older subjects.
Conclusion: IOP measurement by all four methods is affected by CCT. The NCT is affected by CCT significantly more than the GAT. Subject age has a differential effect on the IOP measurements made by the GAT and OBF compared to the Tono-Pen.
Figures


Similar articles
-
A comparison of four methods of tonometry: method agreement and interobserver variability.Br J Ophthalmol. 2005 Jul;89(7):847-50. doi: 10.1136/bjo.2004.056614. Br J Ophthalmol. 2005. PMID: 15965164 Free PMC article. Clinical Trial.
-
The effect of corneal thickness on intraocular pressure measurement in patients with corneal pathology.Br J Ophthalmol. 2004 Nov;88(11):1395-9. doi: 10.1136/bjo.2003.037887. Br J Ophthalmol. 2004. PMID: 15489480 Free PMC article. Clinical Trial.
-
Effect of corneal thickness on intraocular pressure measurements with the pneumotonometer, Goldmann applanation tonometer, and Tono-Pen.Invest Ophthalmol Vis Sci. 2002 May;43(5):1389-92. Invest Ophthalmol Vis Sci. 2002. PMID: 11980851 Clinical Trial.
-
The influence of altitude on the differences between Goldmann tonometry and Pascal dynamic contour tonometry: An ecological meta-analysis.Indian J Ophthalmol. 2024 May 1;72(Suppl 3):S398-S403. doi: 10.4103/IJO.IJO_907_23. Epub 2023 Dec 15. Indian J Ophthalmol. 2024. PMID: 38099358 Free PMC article. Review.
-
Central corneal thickness in glaucoma.Curr Opin Ophthalmol. 2017 Mar;28(2):120-126. doi: 10.1097/ICU.0000000000000335. Curr Opin Ophthalmol. 2017. PMID: 27764022 Review.
Cited by
-
Possibilities of monitoring intraocular pressure in children using EASYTON transpalpebral tonometer.Int Ophthalmol. 2022 May;42(5):1631-1638. doi: 10.1007/s10792-021-02158-5. Epub 2022 Jan 28. Int Ophthalmol. 2022. PMID: 35088357 Free PMC article.
-
Comparison of current tonometry techniques in measurement of intraocular pressure.J Curr Ophthalmol. 2016 Nov 28;29(2):92-97. doi: 10.1016/j.joco.2016.08.010. eCollection 2017 Jun. J Curr Ophthalmol. 2016. PMID: 28626817 Free PMC article.
-
Icare Home Tonometer: A Review of Characteristics and Clinical Utility.Clin Ophthalmol. 2020 Nov 23;14:4031-4045. doi: 10.2147/OPTH.S284844. eCollection 2020. Clin Ophthalmol. 2020. PMID: 33262568 Free PMC article. Review.
-
Correction methods for noncontact intraocular pressure measurement in patients with keratoconus and healthy individuals.Arq Bras Oftalmol. 2022 Sep-Oct;85(5):490-497. doi: 10.5935/0004-2749.20220072. Arq Bras Oftalmol. 2022. PMID: 35170629 Free PMC article.
-
Comparison of intraocular pressure measurements with the tono-pen, goldmann applanation tonometer, and noncontact tonometer in nonglaucomatous pseudophakic children.Int Ophthalmol. 2024 Jun 27;44(1):285. doi: 10.1007/s10792-024-03210-w. Int Ophthalmol. 2024. PMID: 38935310
References
-
- Ehlers N, Bramsen T, Sperling S. Applanation tonometry and central corneal thickness. Acta Ophthalmol 1975;53:34. - PubMed
-
- Whitacre MM, Stein RA, Hassanein K. The effect of corneal thickness on applanation tonometry. Am J Ophthalmol 1993;115:592–6. - PubMed
-
- Gordon MO, Beiser JA, Brandt JD, et al. The ocular hypertension treatment study: baseline factors that predict the onset of primary open-angle glaucoma. Arch Ophthalmol 2002;120:714–20. - PubMed
-
- Dohadwala AA, Munger R, Damji KF. Positive correlation between Tono-Pen intraocular pressure and central corneal thickness. Ophthalmology 1998;105:1849–54. - PubMed
-
- Bhan A, Browning AC, Shah S, et al. Effect of corneal thickness on intraocular pressure measurements with the pneumotonometer, Goldmann applanation tonometer, and Tono-Pen. Invest Ophthalmol Vis Sci 2002;43:1389–92. - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical