Micro ionization chamber dosimetry in IMRT verification: clinical implications of dosimetric errors in the PTV
- PMID: 15967525
- DOI: 10.1016/j.radonc.2005.04.011
Micro ionization chamber dosimetry in IMRT verification: clinical implications of dosimetric errors in the PTV
Abstract
Background and purpose: Absolute dose measurements for Intensity Modulated Radiotherapy (IMRT) beamlets is difficult due to the lack of lateral electron equilibrium. Recently we found that the absolute dosimetry in the penumbra region of the IMRT beamlet, can suffer from significant errors (Capote et al., Med Phys 31 (2004) 2416-2422). This work has the goal to estimate the error made when measuring the Planning Target Volume's (PTV) absolute dose by a micro ion chamber (microIC) in typical IMRT treatment. The dose error comes from the assumption that the dosimetric parameters determining the absolute dose are the same as for the reference conditions.
Materials and methods: Two IMRT treatment plans for common prostate carcinoma case, derived by forward and inverse optimisation, were considered. Detailed geometrical simulation of the microIC and the dose verification set-up was performed. The Monte Carlo (MC) simulation allows us to calculate the delivered dose to water and the dose delivered to the active volume of the ion chamber. However, the measured dose in water is usually derived from chamber readings assuming reference conditions. The MC simulation provides needed correction factors for ion chamber dosimetry in non reference conditions.
Results: Dose calculations were carried out for some representative beamlets, a combination of segments and for the delivered IMRT treatments. We observe that the largest dose errors (i.e. the largest correction factors) correspond to the smaller contribution of the corresponding IMRT beamlets to the total dose delivered in the ionization chamber within PTV.
Conclusion: The clinical impact of the calculated dose error in PTV measured dose was found to be negligible for studied IMRT treatments.
Similar articles
-
Microionization chamber for reference dosimetry in IMRT verification: clinical implications on OAR dosimetric errors.Phys Med Biol. 2005 Mar 7;50(5):959-70. doi: 10.1088/0031-9155/50/5/018. Epub 2005 Feb 17. Phys Med Biol. 2005. PMID: 15798268
-
An EGSnrc Monte Carlo study of the microionization chamber for reference dosimetry of narrow irregular IMRT beamlets.Med Phys. 2004 Sep;31(9):2416-22. doi: 10.1118/1.1767691. Med Phys. 2004. PMID: 15487721
-
Persistently better treatment planning results of intensity-modulated (IMRT) over conformal radiotherapy (3D-CRT) in prostate cancer patients with significant variation of clinical target volume and/or organs-at-risk.Radiother Oncol. 2008 Jul;88(1):77-87. doi: 10.1016/j.radonc.2007.12.011. Epub 2008 Jan 22. Radiother Oncol. 2008. PMID: 18215434
-
Simultaneous beam geometry and intensity map optimization in intensity-modulated radiation therapy.Int J Radiat Oncol Biol Phys. 2006 Jan 1;64(1):301-20. doi: 10.1016/j.ijrobp.2005.08.023. Epub 2005 Nov 14. Int J Radiat Oncol Biol Phys. 2006. PMID: 16289912
-
Practical aspects of inverse-planned intensity-modulated radiation therapy for prostate cancer: a radiation treatment planner's perspective.Can J Urol. 2005 Jun;12 Suppl 2:48-52. Can J Urol. 2005. PMID: 16018834 Review.
Cited by
-
Optimized point dose measurement: An effective tool for QA in intensity-modulated radiotherapy.J Med Phys. 2007 Oct;32(4):156-60. doi: 10.4103/0971-6203.37480. J Med Phys. 2007. PMID: 21224925 Free PMC article.
-
aSi-EPID transit signal calibration for dynamic beams: a needful step for the IMRT in vivo dosimetry.Med Biol Eng Comput. 2013 Oct;51(10):1137-45. doi: 10.1007/s11517-013-1094-x. Epub 2013 Jul 9. Med Biol Eng Comput. 2013. PMID: 23835663 Free PMC article.
-
Characterization of cylindrical ionization chambers for patient specific IMRT QA.J Appl Clin Med Phys. 2009 Sep 30;10(4):241-251. doi: 10.1120/jacmp.v10i4.2923. J Appl Clin Med Phys. 2009. PMID: 19918222 Free PMC article.
-
Statistical analysis of IMRT dosimetry quality assurance measurements for local delivery guideline.Radiat Oncol. 2011 Mar 28;6:27. doi: 10.1186/1748-717X-6-27. Radiat Oncol. 2011. PMID: 21439096 Free PMC article.
-
Optimized point dose measurement for monitor unit verification in intensity modulated radiation therapy using 6 MV photons by three different methodologies with different detector-phantom combinations: A comparative study.J Med Phys. 2010 Jul;35(3):144-50. doi: 10.4103/0971-6203.62129. J Med Phys. 2010. PMID: 20927221 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical