Intravascular ultrasonic assessment of stent diameters derived from manufacturer's compliance charts
- PMID: 15979438
- DOI: 10.1016/j.amjcard.2005.02.049
Intravascular ultrasonic assessment of stent diameters derived from manufacturer's compliance charts
Abstract
We used intravascular ultrasound (IVUS) to assess the accuracy of manufacturers' stent balloon compliance charts. Many interventional cardiologists rely on manufacturers' compliance charts to select stent size and optimize stent diameters according to inflation pressures during percutaneous procedures. We randomly selected 212 patients who had de novo coronary lesions that had been treated with a single, bare metal, > or =3.0-mm stent (Bx velocity, NIR, TETRA/PENTA, S660/S670/S7) under IVUS guidance. Cases of stent overlap and postdilatation with another balloon were excluded. Predicted stent diameters were derived from each manufacturer's compliance charts, and stent size and final maximal deployment pressures were derived from each physician's report. IVUS-measured minimum stent diameters (range 1.4 to 4.0 mm, mean 2.79 +/- 0.48) were smaller than predicted diameters (range 3.1 to 4.57 mm, mean 3.79 +/- 0.44). The ratio of IVUS to predicted diameters ranged from 44% to 97% (mean 74 +/- 10%). This finding was common to all 3 stent sizes: 74 +/- 12% for 3.0 mm, 73 +/- 9% for 3.5 mm, and 74 +/- 9% for 4.0-mm stents (p = 0.9). This finding was also common to all 4 stent manufacturers, 72 +/- 8% for Boston Scientific, 76 +/- 11% for Guidant, 73 +/- 9% for Cordis, and 74 +/- 11% for Medtronic (p = 0.13), and to different stent lengths. Only 3.8% of the stents achieved 90% of the predicted minimum stent diameters, and only 24.6% achieved 80% of the predicted minimum stent diameters. In conclusion, in human coronary arteries, minimal stent diameter measured by IVUS is significantly smaller than that predicted by in vitro compliance charts. These differences are independent of stent manufacturer, length, diameter, and deployment pressure.
Similar articles
-
Is adjunctive balloon postdilatation necessary after coronary stent deployment? Final results from the POSTIT trial.Catheter Cardiovasc Interv. 2003 Jun;59(2):184-92. doi: 10.1002/ccd.10474. Catheter Cardiovasc Interv. 2003. PMID: 12772236 Clinical Trial.
-
Intravascular ultrasound assessment of drug-eluting stent expansion.Am Heart J. 2007 Feb;153(2):297-303. doi: 10.1016/j.ahj.2006.08.026. Am Heart J. 2007. PMID: 17239693
-
Comparison of myocardial fractional flow reserve and intravascular ultrasound for the assessment of slotted-tube stents.Catheter Cardiovasc Interv. 2001 Mar;52(3):322-6; discussion 327. doi: 10.1002/ccd.1074. Catheter Cardiovasc Interv. 2001. PMID: 11246245
-
Adjunctive balloon postdilatation after stent deployment: is it still necessary with drug-eluting stents?J Interv Cardiol. 2006 Feb;19(1):43-50. doi: 10.1111/j.1540-8183.2006.00103.x. J Interv Cardiol. 2006. PMID: 16483339 Review.
-
Digital enhancement of stent images in primary and secondary percutaneous coronary revascularisation.EuroIntervention. 2009 May;5 Suppl D:D101-6. EuroIntervention. 2009. PMID: 19736057 Review.
Cited by
-
Effect of stent inflation pressure and post-dilatation on the outcome of coronary artery intervention. A report of more than 90,000 stent implantations.PLoS One. 2013;8(2):e56348. doi: 10.1371/journal.pone.0056348. Epub 2013 Feb 13. PLoS One. 2013. PMID: 23418560 Free PMC article.
-
Compliance charts to guide non-complex small artery stenting: validation by quantitative coronary angiography.Heart Asia. 2013 May 31;5(1):76-9. doi: 10.1136/heartasia-2013-010312. eCollection 2013. Heart Asia. 2013. PMID: 27326085 Free PMC article.
-
Expansion of the Multi-Link Frontier™ coronary bifurcation stent: micro-computed tomographic assessment in human autopsy and porcine heart samples.PLoS One. 2011;6(7):e21778. doi: 10.1371/journal.pone.0021778. Epub 2011 Jul 21. PLoS One. 2011. PMID: 21814552 Free PMC article.
-
Imaging during percutaneous coronary intervention for optimizing outcomes.Indian Heart J. 2018 Dec;70 Suppl 3(Suppl 3):S456-S465. doi: 10.1016/j.ihj.2018.08.012. Epub 2018 Aug 18. Indian Heart J. 2018. PMID: 30595307 Free PMC article. Review.
-
Two-in-one aortic valve sizing and valvuloplasty conductance balloon catheter.Catheter Cardiovasc Interv. 2015 Jul;86(1):136-43. doi: 10.1002/ccd.25774. Epub 2015 Jan 30. Catheter Cardiovasc Interv. 2015. PMID: 25510238 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous