Evaluation of a hot-wire respiratory flowmeter for clinical applicability
- PMID: 159886
- DOI: 10.1152/jappl.1979.47.5.1131
Evaluation of a hot-wire respiratory flowmeter for clinical applicability
Abstract
A hot-wire flowmeter was evaluated for clinical applicability. 1) Calibration with a syringe could be done if emptying duration was 1-15 s. 2) Outputs linearly decreased with decreasing concentration of N2 in O2. Outputs with CO2 in O2 and N2O in O2 represented nonlinear convex relationships with varying concentrations of CO2 and N2O. Correction for each gas mixture to be measured is required. 3) Outputs linearly decreased with decreasing barometric pressure. 4) Stability assessed with a piston respirator was excellent (within +/- 2% of syringe volume) after 15 min warmup time. However, daily calibrations are recommended in clinical situations. 5) Nebulization, if not excessive, was acceptable if the expired gas was measured at the mouth. 6) Hot-wire burning, which occurred when it was partially in contact with materials whose specific heat differed with air, was successfully protected with a simple shutoff circuit. 7) The possibility of producing nitrogen oxides by the catalytic action of the platinum hot-wire was denied by colorimetric determination. Interchangeability and sterilizability of transducers and improved mechanical strength with platinum-rhodium alloy are also discussed.
Similar articles
-
Effects of lung volume and O2 and CO2 content on cutaneous gas exchange in frogs.Am J Physiol. 1986 Nov;251(5 Pt 2):R941-6. doi: 10.1152/ajpregu.1986.251.5.R941. Am J Physiol. 1986. PMID: 3096154
-
Fresh gas flow and carbon dioxide rebreathing in a low pressure semi-open anaesthesia system.Can J Anaesth. 1993 Nov;40(11):1096-101. doi: 10.1007/BF03009484. Can J Anaesth. 1993. PMID: 8269574
-
A bidirectional respiratory flowmeter using the hot-wire principle.J Appl Physiol. 1975 Feb;38(2):360-5. doi: 10.1152/jappl.1975.38.2.360. J Appl Physiol. 1975. PMID: 1120766
-
Expirograms of O2, CO2 and intravenously infused C2H2 and Freon-22 during panting in dogs.Respir Physiol. 1990 May-Jun;80(2-3):171-9. doi: 10.1016/0034-5687(90)90081-9. Respir Physiol. 1990. PMID: 2120751
-
Flow measurement in mechanical ventilation: a review.Med Eng Phys. 2015 Mar;37(3):257-64. doi: 10.1016/j.medengphy.2015.01.010. Epub 2015 Feb 7. Med Eng Phys. 2015. PMID: 25659299 Review.
Cited by
-
A novel, non-invasive method of respiratory monitoring for use with stereotactic procedures.J Neurosci Methods. 2012 Aug 15;209(2):337-43. doi: 10.1016/j.jneumeth.2012.06.029. Epub 2012 Jul 4. J Neurosci Methods. 2012. PMID: 22771713 Free PMC article.
-
Effects of caudal block with mepivacaine on resting ventilation and ventilatory response to carbon dioxide in sedated children.Can J Anaesth. 1988 Jul;35(4):354-8. doi: 10.1007/BF03010855. Can J Anaesth. 1988. PMID: 3135953
-
ERS statement on standardisation of cardiopulmonary exercise testing in chronic lung diseases.Eur Respir Rev. 2019 Dec 18;28(154):180101. doi: 10.1183/16000617.0101-2018. Print 2019 Dec 31. Eur Respir Rev. 2019. PMID: 31852745 Free PMC article. Review.
-
Functional residual capacity measurement during tracheal gas insufflation.J Clin Monit Comput. 1998 May;14(4):225-32. doi: 10.1023/a:1009998100819. J Clin Monit Comput. 1998. PMID: 9754610
-
A novel sensor for routine continuous spirometry of intubated patients.J Clin Monit. 1993 Nov;9(5):374-80. doi: 10.1007/BF01618680. J Clin Monit. 1993. PMID: 8106891