. 2005 Jun;58(6):1194-201.
doi: 10.1097/01.ta.0000170400.68994.ab.

Locked plate fixation and intramedullary nailing for proximal humerus fractures: a biomechanical evaluation

Affiliations

Locked plate fixation and intramedullary nailing for proximal humerus fractures: a biomechanical evaluation

Martin Henri Hessmann et al. J Trauma. 2005 Jun.

Abstract

Background: Internal fixation of proximal humeral fractures is associated with a considerable secondary malalignment rate. Fixed-angle implants have been suggested to increase the stability of fixation.

Methods: The biomechanical properties of four different implants were tested. These included an internal fixator with semi-elastic properties (reference), the Synthes T-plate, a locked plate with rigid properties and a spiral blade locked intramedullary nail (PHN). These implants were assessed in twenty-four osteotomized pairs of human cadaveric humeri. Specimens were subjected to two-hundred cycles of axial loading and torque followed by load to failure.

Results: The PHN had greater axial stiffness than the reference and the T-plate. During torque, all implants were stiffer than the reference. During cyclic loading, there were no differences between the T-plate and the reference. Both the rigid internal fixator and the PHN had less irreversible deformation than the reference. Both implants resisted higher loads before failure.

Conclusion: This study showed that the proximal humeral nail and the rigid internal fixator are stronger than the semi-elastic locked plate and the Synthes T-plate for unstable subcapital proximal humeral fractures.