Value of intraoperative true lateral radiograph of C2 pedicle for C1-2 transarticular screw insertion
- PMID: 15996613
- DOI: 10.1016/j.spinee.2005.01.009
Value of intraoperative true lateral radiograph of C2 pedicle for C1-2 transarticular screw insertion
Abstract
Background context: Transarticular C1-2 screws are widely used in posterior cervical spine instrumentation. Injury to the vertebral artery during insertion of transarticular Cl-2 screw remains a serious complication. Use of a computer-assisted surgery system decreases this complication considerably. However, this system encounters problems in ensuring complete accuracy because of positional variations during preoperative and intraoperative imaging generation. Therefore, intraoperative fluoroscopy still is one of the commonly used methods to guide insertion of transarticular Cl-2 screw. Evaluation of a true lateral radiographic view of the C2 pedicle for screw trajectory during C1-2 transarticular screw insertion may help to minimize this potential complication.
Purpose: To evaluate the value of intraoperative true lateral radiograph of the C2 pedicle for screw trajectory during C1-2 transarticular screw insertion.
Study design: To compare the height of the C2 pedicle area allowing instrumentation on true lateral view radiograph of the C2 pedicle and computed tomographic (CT) scan with multiplanar reconstruction.
Methods: Twenty embalmed human cadaveric cervical spine specimens were used to insert a total of 40 C1-2 transarticular screws using Magerl and Seemann technique. One side of the C2 transverse foramen was filled with radiopaque material (lead oxide) to simulate the artery and to demarcate the danger zone for better visualization on radiography. Measurements and calculation of the mean and standard deviation of the height of the area allowing instrumentation of the C2 pedicle were done on true lateral view radiograph of the C2 pedicle, the sagittal and 30 degrees sagittal views relative to the frontal plane passing exactly through the center of the C2 pedicle of CT scans. Student t test was applied to calculate the statistical significance of measured values. Statistical significance was defined as p<or=.001.
Results: On true lateral radiographic views of the C2 pedicle, the height of the area allowing instrumentation of the pedicle was 7.75+/-0.92 mm (right) and 7.64+/-0.63 (left), p>or=.36. Using sagittal CT scan views, the height of pedicles was 7.71+/-0.7 mm (right) and 7.58+/-1.01 mm (left), p>or=.23. On 30 degrees sagittal CT scan views, the height of pedicles was 7.84+/-1.00 mm (right) and 7.76+/-1.02 mm (left), p>or=.27. The p value was >or=.78, >or=.56, and >or=.49 for true lateral radiographic view and sagittal CT scan view, true lateral radiographic view and 30 degrees sagittal CT scan view, and sagittal CT scan view and 30 degrees sagittal CT scan views, respectively. On lateral view of cervical spine, the decline angle of the transarticular screw was 51.3+/-0.50 degrees (right) and 50.68+/-0.41 degrees (left), p>or=.17. Mean decline angle was 51+/-0.43 degrees . On the anteroposterior (AP) view, radiograph median angle was 6.87+/-0.53 degrees (right) and 6.0+/-0.59 degrees (left), p>or=.25. Mean median angle was 6.44+/-0.62 degrees.
Conclusions: True lateral radiographic views of the pedicles provide useful information for defining screw trajectory intraoperatively. Using this view along with AP and lateral view of cervical spine and preoperative three-dimensional CT scan may narrow the margin of error in this delicate area.
Similar articles
-
Risk of vertebral artery injury: comparison between C1-C2 transarticular and C2 pedicle screws.Spine J. 2013 Jul;13(7):775-85. doi: 10.1016/j.spinee.2013.04.005. Epub 2013 May 14. Spine J. 2013. PMID: 23684237
-
Anatomical considerations for the placement of cervical transarticular screws.J Neurosurg Spine. 2011 Jan;14(1):114-21. doi: 10.3171/2010.9.SPINE1066. Epub 2010 Dec 17. J Neurosurg Spine. 2011. PMID: 21166487
-
Development of a new technique for pedicle screw and Magerl screw insertion using a 3-dimensional image guide.Spine (Phila Pa 1976). 2012 Nov 1;37(23):1983-8. doi: 10.1097/BRS.0b013e31825ab547. Spine (Phila Pa 1976). 2012. PMID: 22531473
-
Potential intraoperative factors of screw-related complications following posterior transarticular C1-C2 fixation: a systematic review and meta-analysis.Eur Spine J. 2019 Feb;28(2):400-420. doi: 10.1007/s00586-018-5830-7. Epub 2018 Nov 22. Eur Spine J. 2019. PMID: 30467736
-
Comparison of two novel fluoroscopy-based stereotactic methods for cervical pedicle screw placement and review of the literature.Eur Spine J. 2008 Apr;17(4):564-75. doi: 10.1007/s00586-008-0584-2. Epub 2008 Jan 22. Eur Spine J. 2008. PMID: 18210169 Free PMC article. Review.